Recalls and safety alerts involving pacemakers and implantable cardioverter-defibrillator generators
- PMID: 11497532
- DOI: 10.1001/jama.286.7.793
Recalls and safety alerts involving pacemakers and implantable cardioverter-defibrillator generators
Abstract
Context: Unanticipated pacemaker and implantable cardioverter-defibrillator (ICD) generator malfunctions sometimes warrant recall by the US Food and Drug Administration (FDA). Despite increasingly frequent device implantation, pacemaker and ICD recalls and safety alerts (advisories) remain poorly characterized.
Objectives: To determine pacemaker and ICD generator advisory rates in the United States, to identify trends in these rates, and to examine their clinical and financial implications.
Design and setting: Analysis of weekly FDA Enforcement Reports issued between January 1990 and December 2000 to identify all advisories involving pacemaker or ICD generators in the United States. Recalls and safety alerts involving lead malfunctions were not included.
Main outcome measures: Number of pacemakers and ICD generators in the United States subject to FDA recall or safety alert in 1990-2000; annual pacemaker and ICD advisory rates in the United States in 1990-2000; and estimated cost of device advisories.
Results: During the study period, 52 advisories (median [25th and 75th percentiles], 4 [4 and 7] per year) involving 408 500 pacemakers and 114 645 ICDs (523 145 total devices) were issued. Hardware malfunctions (35 advisories affecting 280 641 devices) and computer errors (10 advisories affecting 216 533 devices) accounted for 95% of device recalls. Implantable cardioverter-defibrillators were recalled more frequently than pacemakers (mean [SD], 16.4 [1.6] vs 6.7 [0.8] advisories per 100 person-years; P<.001). Between 1995 and 2000, the annual advisory rate increased for both pacemakers (P for trend <.001) and ICDs (P for trend =.02). An estimated 1.3 million device checks and analyses and 36 187 device replacements resulted from the advisories and cost approximately $870 million.
Conclusions: Pacemaker and ICD recalls and safety alerts occur frequently, affect many patients, and appear to be increasing in number and rate. With the growing number of device implants and expanding indications for device therapy, the number of patients affected by device advisories will likely continue to increase.
Comment in
-
Safety alerts involving device therapy for arrhythmias.JAMA. 2001 Aug 15;286(7):843-4. doi: 10.1001/jama.286.7.843. JAMA. 2001. PMID: 11497539 No abstract available.
-
Safety of implantable cardiac devices.JAMA. 2002 Feb 20;287(7):839-40. doi: 10.1001/jama.287.7.839. JAMA. 2002. PMID: 11851565 No abstract available.
-
High quality performance of pacemakers and implantable defibrillators.Pacing Clin Electrophysiol. 2002 Dec;25(12):1667-9. doi: 10.1046/j.1460-9592.2002.01667.x. Pacing Clin Electrophysiol. 2002. PMID: 12520665 No abstract available.
Similar articles
-
Changing trends in pacemaker and implantable cardioverter defibrillator generator advisories.Pacing Clin Electrophysiol. 2002 Dec;25(12):1670-8. doi: 10.1046/j.1460-9592.2002.01670.x. Pacing Clin Electrophysiol. 2002. PMID: 12520666
-
Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports.JAMA. 2006 Apr 26;295(16):1901-6. doi: 10.1001/jama.295.16.1901. JAMA. 2006. PMID: 16639048
-
Recalls and safety alerts affecting automated external defibrillators.JAMA. 2006 Aug 9;296(6):655-60. doi: 10.1001/jama.296.6.655. JAMA. 2006. PMID: 16896108
-
Should recent defibrillator and lead advisories affect decisions to refer patients for implantable cardioverter-defibrillator therapy?Curr Opin Cardiol. 2010 Jan;25(1):23-8. doi: 10.1097/HCO.0b013e328333d375. Curr Opin Cardiol. 2010. PMID: 19907318 Review.
-
[Postmarketing surveillance in patients with cardiac pace-makers or automatic implantable defibrillators].Ann Cardiol Angeiol (Paris). 2005 Jan;54(1):38-43. doi: 10.1016/j.ancard.2004.11.004. Ann Cardiol Angeiol (Paris). 2005. PMID: 15702910 Review. French.
Cited by
-
Canadian Registry of Electronic Device Outcomes (CREDO): The Abbott ICD Premature Battery Depletion Advisory, a Multicentre Cohort Study.CJC Open. 2020 Sep 12;3(1):48-53. doi: 10.1016/j.cjco.2020.09.008. eCollection 2021 Jan. CJC Open. 2020. PMID: 33458632 Free PMC article.
-
Fully Bioabsorbable Capacitor as an Energy Storage Unit for Implantable Medical Electronics.Adv Sci (Weinh). 2019 Jan 22;6(6):1801625. doi: 10.1002/advs.201801625. eCollection 2019 Mar 20. Adv Sci (Weinh). 2019. PMID: 30937259 Free PMC article.
-
Magnetic resonance imaging, pacemakers and implantable cardioverter-defibrillators: current situation and clinical perspective.Neth Heart J. 2010 Jan;18(1):31-7. Neth Heart J. 2010. PMID: 20111641 Free PMC article.
-
Therapy for ventricular arrhythmias in structural heart disease: a multifaceted challenge.J Physiol. 2016 May 1;594(9):2431-43. doi: 10.1113/JP270534. Epub 2016 Jan 27. J Physiol. 2016. PMID: 26621333 Free PMC article. Review.
-
"Add-on pacing lead": an effective and safe alternative to lead replacement in ICD pacing failure.Indian Heart J. 2012 Jul-Aug;64(4):420-2. doi: 10.1016/j.ihj.2012.06.012. Epub 2012 Jun 27. Indian Heart J. 2012. PMID: 22929831 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical