Characterization of the 16 blanking periods of the Medtronic GEM DR dual chamber defibrillators
- PMID: 11500587
- DOI: 10.1023/a:1011477002235
Characterization of the 16 blanking periods of the Medtronic GEM DR dual chamber defibrillators
Abstract
We determined the blanking periods of the Medtronic GEM DR dual chamber defibrillators by using a simulator to deliver signals mimicking arrhythmias into external devices programmed to various settings. The blanking periods for the tachycardia and bradycardia functions were measured in the atrial and ventricular channels after a paced atrial event, sensed atrial event, paced ventricular event and sensed ventricular event, adding to a total of 16 blanking periods. Our findings complement the incomplete or unclear specifications published by the manufacturer. Accurate knowledge of blanking periods is essential for the interpretation of device function.
Similar articles
-
Effect of the atrial blanking time on the detection of atrial fibrillation in dual chamber pacing.Pacing Clin Electrophysiol. 2001 Apr;24(4 Pt 1):496-9. doi: 10.1046/j.1460-9592.2001.00496.x. Pacing Clin Electrophysiol. 2001. PMID: 11341088
-
Sensing and detection in Medtronic implantable cardioverter defibrillators.Herzschrittmacherther Elektrophysiol. 2016 Sep;27(3):193-212. doi: 10.1007/s00399-016-0450-6. Epub 2016 Sep 8. Herzschrittmacherther Elektrophysiol. 2016. PMID: 27624809 Review. English.
-
Inappropriate shock therapy for nonsustained ventricular tachycardia in a dual chamber pacemaker defibrillator.Pacing Clin Electrophysiol. 1998 Oct;21(10):2005-6. doi: 10.1111/j.1540-8159.1998.tb00024.x. Pacing Clin Electrophysiol. 1998. PMID: 9793101
-
Worldwide clinical experience with a new dual-chamber implantable cardioverter defibrillator system.J Cardiovasc Electrophysiol. 2001 May;12(5):521-8. doi: 10.1046/j.1540-8167.2001.00521.x. J Cardiovasc Electrophysiol. 2001. PMID: 11386511
-
Supraventricular tachycardia-ventricular tachycardia discrimination algorithms in implantable cardioverter defibrillators: state-of-the-art review.J Cardiovasc Electrophysiol. 2001 May;12(5):606-12. doi: 10.1046/j.1540-8167.2001.00606.x. J Cardiovasc Electrophysiol. 2001. PMID: 11386525 Review.
Cited by
-
Ventricular rate stabilization algorithm of ICD causing dual chamber pacing during ventricular tachycardia.J Interv Card Electrophysiol. 2003 Dec;9(3):397-400. doi: 10.1023/a:1027412014937. J Interv Card Electrophysiol. 2003. PMID: 14618063
-
A reappraisal of pacemaker timing cycles pertaining to automatic mode switching.J Interv Card Electrophysiol. 2001 Dec;5(4):417-29. doi: 10.1023/a:1013298012367. J Interv Card Electrophysiol. 2001. PMID: 11752910 Review.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials