The Clinac-18 linear accelerator. 10 MeV photon beam data
- PMID: 115120
The Clinac-18 linear accelerator. 10 MeV photon beam data
Abstract
Measurements were carried out on a Clinac-18 linear accelerator in order to characterize the 10 MeV photon beam. Depth dose data and tissue-maximum-ratios were obtained for field sizes up to 35 x 35 cm2. Using a parabolic expression a least square fit to the tissue-maximum-ratios was performed. The radiation field flatness and the penumbra were found to be in good agreement with the tolerances recommended by the IEC standard (draft). The radiation field and the light field isocentre were determined as suggested by the IEC. The distance between both isocentres was less than 1 mm.
Similar articles
-
Optimization of photon beam flatness for radiation therapy.Phys Med Biol. 2007 Mar 21;52(6):1735-46. doi: 10.1088/0031-9155/52/6/013. Epub 2007 Feb 27. Phys Med Biol. 2007. PMID: 17327659
-
Characterization of megavoltage electron beams delivered through a photon multi-leaf collimator (pMLC).Phys Med Biol. 2006 Apr 21;51(8):2113-29. doi: 10.1088/0031-9155/51/8/011. Epub 2006 Apr 3. Phys Med Biol. 2006. PMID: 16585849
-
[Measurement and calculation of the air activation caused by 15 MeV continuous radiation emitted by a linear accelerator (author's transl)].Strahlentherapie. 1981 Mar;157(3):187-90. Strahlentherapie. 1981. PMID: 6782714 German.
-
Development and commissioning of a Monte Carlo photon beam model for Varian Clinac 2100EX linear accelerator.Appl Radiat Isot. 2006 Jun;64(6):656-62. doi: 10.1016/j.apradiso.2005.12.012. Epub 2006 Feb 7. Appl Radiat Isot. 2006. PMID: 16455264
-
Beam configuration and physical parameters of clinical high energy photon beam for total body irradiation (TBI).Phys Med. 2011 Jul;27(3):163-8. doi: 10.1016/j.ejmp.2010.10.003. Epub 2010 Nov 11. Phys Med. 2011. PMID: 21074467