A novel low wearing differential hardness, ceramic-on-metal hip joint prosthesis
Abstract
Osteolysis and loosening of artificial joints caused by polyethylene wear debris has prompted renewed interest in alternative bearing materials for hip prosthesis designs. Lower wearing metal-on-metal (MOM) and ceramic-on-ceramic prostheses are being used more extensively, and there is considerable interest in further improving on their performance. This study investigated the wear properties and debris morphology of a novel differential hardness ceramic-on-metal (COM) prosthesis, in comparison with MOM articulations in a physiological anatomical hip joint simulator. The COM pairings were found to have wear rates approximately 100-fold lower than the MOM pairings. The MOM pairings showed a higher "bedding in" wear rate (3.09+/-0.46mm(3)/10(6) cycles) in the first million cycles, which then reduced to a steady state wear rate of 1.23+/-0.5mm(3)/10(6) cycles. The wear rate of the COM pairings over the duration of the test was approximately 0.01mm(3)/10(6) cycles with very little wear detected on the surface of the prosthesis components. The wear particles from both articulations were oval to round in shape and in the nanometer size range. After one million cycles the mean maximum diameter of the MOM and COM wear particles were 30+/-2.25 and 17.57+/-1.37nm, respectively. After five million cycles the wear particles were statistically significantly smaller than at one million cycles, 13.9+/-0.72nm for the MOM pairings and 6.11+/-0.40nm for the COM pairings. The wear rates of the MOM prostheses were representative of clinical values. The use of differential hardness COM pairings dramatically reduced the wear rate compared to MOM hip prostheses. The wear particles from the MOM articulation were similar to particles found in retrieved tissues from around MOM prostheses. The extremely low wearing differential hardness COM bearings presented in this study produced far smaller volumetric particle loads compared to MOM prostheses currently used clinically.
Similar articles
-
Comparison of ceramic-on-metal and metal-on-metal hip prostheses under adverse conditions.J Biomed Mater Res B Appl Biomater. 2013 Jul;101(5):770-5. doi: 10.1002/jbm.b.32880. Epub 2013 Jan 29. J Biomed Mater Res B Appl Biomater. 2013. PMID: 23359608
-
Quantitative analysis of wear and wear debris from metal-on-metal hip prostheses tested in a physiological hip joint simulator.Biomed Mater Eng. 2001;11(2):143-57. Biomed Mater Eng. 2001. PMID: 11352113
-
A comparative joint simulator study of the wear of metal-on-metal and alternative material combinations in hip replacements.Proc Inst Mech Eng H. 2000;214(1):39-47. doi: 10.1243/0954411001535228. Proc Inst Mech Eng H. 2000. PMID: 10718049
-
Biological reactions to wear debris in total joint replacement.Proc Inst Mech Eng H. 2000;214(1):21-37. doi: 10.1243/0954411001535219. Proc Inst Mech Eng H. 2000. PMID: 10718048 Review.
-
Hard bearing surfaces in total hip arthroplasty.Am J Orthop (Belle Mead NJ). 1998 Apr;27(4):283-92. Am J Orthop (Belle Mead NJ). 1998. PMID: 9586727 Review.
Cited by
-
A comparison of metal/metal and ceramic/metal taper-trunnion modular connections in explanted total hip replacements.J Biomed Mater Res B Appl Biomater. 2022 Jan;110(1):135-143. doi: 10.1002/jbm.b.34897. Epub 2021 Jun 23. J Biomed Mater Res B Appl Biomater. 2022. PMID: 34164932 Free PMC article.
-
Adverse reaction to metal debris in a patient with acetabular shell loosening 8 years after ceramic-on-metal total hip arthroplasty.Arthroplast Today. 2015 Oct 21;1(4):93-98. doi: 10.1016/j.artd.2015.07.002. eCollection 2015 Dec. Arthroplast Today. 2015. PMID: 28326381 Free PMC article.
-
No Difference in Reoperations at 2 Years Between Ceramic-on-metal and Metal-on-metal THA: A Randomized Trial.Clin Orthop Relat Res. 2016 Feb;474(2):447-55. doi: 10.1007/s11999-015-4424-5. Clin Orthop Relat Res. 2016. PMID: 26135473 Free PMC article. Clinical Trial.
-
Influence of Metallic Deposition on Ceramic Femoral Heads on the Wear Behavior of Artificial Hip Joints: A Simulator Study.Materials (Basel). 2020 Aug 12;13(16):3569. doi: 10.3390/ma13163569. Materials (Basel). 2020. PMID: 32806756 Free PMC article.
-
The John Charnley Award: an accurate and sensitive method to separate, display, and characterize wear debris: part 1: polyethylene particles.Clin Orthop Relat Res. 2012 Feb;470(2):329-38. doi: 10.1007/s11999-011-2057-x. Clin Orthop Relat Res. 2012. PMID: 21997780 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources