. 2001 Sep;142(3):482-8.
doi: 10.1067/mhj.2001.116770.

The role of transesophageal echocardiography in transcatheter closure of secundum atrial septal defects by the Amplatzer septal occluder

Affiliations

The role of transesophageal echocardiography in transcatheter closure of secundum atrial septal defects by the Amplatzer septal occluder

U Mazic et al. Am Heart J. 2001 Sep.

Abstract

Background: Our purpose was to determine the role of transesophageal echocardiography (TEE) in the closure of atrial septal defects by the Amplatzer septal occluder (ASO) (AGA Medical, Golden Valley, Minn).

Methods: A total of 240 patients with atrial septal defect (ASD) secundum were examined by transthoracic 2-dimensional echocardiography (TTE) and TEE to determine the ASD morphologic features, diameter, and rims. During transcatheter closure TEE was used for determination of the ASD diameter and guidance of the ASO implantation.

Results: Sixteen (6%) patients were found not suitable for transcatheter closure with TTE, 35 (14%) with TEE, and 2 during catheterization. Twenty-eight patients (18%) had partial or total deficiency of the posterior, inferoanterior, or inferoposterior rim, 54 (27%) had a centrally positioned ASD, 92 (46%) had insufficient superoanterior rim, and 9 had multiple ASDs, whereas 8 had a septal aneurysm associated with a single defect and 4 a multiperforated aneurysm. A total of 170 patients underwent implantation of ASO. The ASO was correctly positioned in 144 at the first attempt. In the remainder TEE revealed unstable position of the left atrial disk (12), opening of both atrial disks in the left atrium (5), deployment of the device through the smaller defect in patients with multiple ASDs (3), and, in 1 patient, the device was too small and had to be replaced by a larger one.

Conclusions: Morphologic variations of the ASD are common. TEE is crucial for the determination of the ASD morphologic features, diameter, and rims, which are crucial for proper patient selection. TEE allows precise guiding and positioning of the ASO, which is essential for safe and effective transcatheter ASD closure.

PubMed Disclaimer

Publication types

LinkOut - more resources