Estimation of end-expiratory lung volume variations by optoelectronic plethysmography
- PMID: 11546992
- DOI: 10.1097/00003246-200109000-00026
Estimation of end-expiratory lung volume variations by optoelectronic plethysmography
Abstract
Objective: To test the capability of optoelectronic plethysmography (OEP) to monitor positive end-expiratory pressure (PEEP)-induced changes of end-expiratory lung volume (EELV) changes in mechanically ventilated paralyzed patients.
Design: Laboratory and clinical investigation.
Setting: Intensive care unit of the Ospedale Maggiore Policlinico di Milano.
Patients: A total of eight patients with respiratory failure of various degrees, sedated and paralyzed.
Interventions: PEEP variations (+/-5 cm H2O) relative to the baseline PEEP of 10 cm H2O.
Measurements and main results: In the model protocol, we tested the reproducibility of the OEP by repeating volume measurements of a plastic torso model over a 21-hr period, every 30 mins. The variations of OEP measurements of the torso model (9337 mL value) were encountered in a range of 16 mL (sd = 4 mL). In the patient protocol, we measured the end-expiratory volume of the chest wall (EEVCW) breath-by-breath by OEP before, during, and after the PEEP increase/decrease and we compared its variations with the corresponding variations of EELV measured by helium dilution technique. The regression line between EELV changes measured by helium and EEVCW changes measured by OEP resulted very close to the identity line (slope 1.06, intercept -0.02 L, r(2) = 0.89) and their difference was not related to their absolute magnitude. After PEEP increase, the new steady state of EEVCW was reached approximately in 15 breaths; and, after PEEP decrease, in 3-4 breaths. The slow increase in EEVCW was mainly because of the abdominal compartment.
Conclusion: OEP measurements of EEVCW accurately reflect the changes of EELV. Furthermore, OEP allows a continuous compartmental analysis, even during unsteady conditions.
Similar articles
-
Does alveolar recruitment occur with positive end-expiratory pressure in adult respiratory distress syndrome patients?J Crit Care. 1993 Mar;8(1):34-42. doi: 10.1016/0883-9441(93)90031-f. J Crit Care. 1993. PMID: 8343857
-
Evaluation of respiratory inductive plethysmography in controlled ventilation: measurement of tidal volume and PEEP-induced changes of end-expiratory lung volume.Chest. 1998 Feb;113(2):443-51. doi: 10.1378/chest.113.2.443. Chest. 1998. PMID: 9498965
-
Evaluation of respiratory inductive plethysmography in the measurement of breathing pattern and PEEP-induced changes in lung volume.Chest. 1992 Jul;102(1):234-8. doi: 10.1378/chest.102.1.234. Chest. 1992. PMID: 1623760
-
Optoelectronic plethysmography in intensive care patients.Am J Respir Crit Care Med. 2000 May;161(5):1546-52. doi: 10.1164/ajrccm.161.5.9903024. Am J Respir Crit Care Med. 2000. PMID: 10806152
-
Use of recruitment maneuvers and high-positive end-expiratory pressure in a patient with acute respiratory distress syndrome.Crit Care Med. 2000 Apr;28(4):1210-6. doi: 10.1097/00003246-200004000-00051. Crit Care Med. 2000. PMID: 10809308 Review.
Cited by
-
Updated Perspectives on the Role of Biomechanics in COPD: Considerations for the Clinician.Int J Chron Obstruct Pulmon Dis. 2022 Oct 17;17:2653-2675. doi: 10.2147/COPD.S339195. eCollection 2022. Int J Chron Obstruct Pulmon Dis. 2022. PMID: 36274993 Free PMC article. Review.
-
Algorithms, modelling and VO₂ kinetics.Eur J Appl Physiol. 2011 Mar;111(3):331-42. doi: 10.1007/s00421-010-1396-8. Epub 2010 Feb 27. Eur J Appl Physiol. 2011. PMID: 20195628 Review.
-
In Vivo Measurement of Tidal Volume During Non-invasive Respiratory Support by Continuous-Flow Helmet CPAP.Ann Biomed Eng. 2024 Sep;52(9):2546-2555. doi: 10.1007/s10439-024-03545-6. Epub 2024 Jun 17. Ann Biomed Eng. 2024. PMID: 38886251 Free PMC article.
-
Correlated variability in the breathing pattern and end-expiratory lung volumes in conscious humans.PLoS One. 2015 Mar 24;10(3):e0116317. doi: 10.1371/journal.pone.0116317. eCollection 2015. PLoS One. 2015. PMID: 25803710 Free PMC article.
-
Open-circuit respirometry: real-time, laboratory-based systems.Eur J Appl Physiol. 2018 May;118(5):875-898. doi: 10.1007/s00421-018-3860-9. Epub 2018 May 4. Eur J Appl Physiol. 2018. PMID: 29728765 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources