Comparative Study
. 2001 Oct;39(10):3633-6.
doi: 10.1128/JCM.39.10.3633-3636.2001.

Comparison of the Vitek gram-positive susceptibility 106 card, the MRSA-Screen latex agglutination test, and mecA analysis for detecting oxacillin resistance in a geographically diverse collection of clinical isolates of coagulase-negative staphylococci

Affiliations
Comparative Study

Comparison of the Vitek gram-positive susceptibility 106 card, the MRSA-Screen latex agglutination test, and mecA analysis for detecting oxacillin resistance in a geographically diverse collection of clinical isolates of coagulase-negative staphylococci

T Yamazumi et al. J Clin Microbiol. 2001 Oct.

Abstract

The Vitek automated susceptibility testing system with a modified gram-positive susceptibility (GPS) 106 card (bioMerieux Vitek, Inc., Hazelwood. Mo.) and a rapid slide latex agglutination test (MRSA-Screen test; Denka Seiken Co., Ltd., Tokyo, Japan) were evaluated for their abilities to detect oxacillin resistance in coagulase-negative staphylococci (CoNS). The reference broth microdilution method and the detection of the mecA gene by PCR ("gold standard" reference result) were used to compare the results obtained with the commercial products. A total of 123 clinical isolates consisting of eight species were selected from U.S. surveillance collections. Among the mecA-positive isolates (95 strains), 30 isolates were initially negative on the MRSA-Screen test read at 3 min. When the agglutination reaction was extended for 10 min, 26 of the 30 isolates became positive. For a different four isolates, the oxacillin MIC was < or =0.25 microg/ml on the Vitek GPS 106 card. Among the mecA-negative isolates (28 strains), for two Staphylococcus warneri, two S. lugdunensis, and two S. saprophyticus strains MICs were > or =0.5 microg/ml by the reference broth microdilution method. Four of these isolates were also categorized as resistant with the Vitek GPS 106 card and two isolates were positive by the MRSA-Screen test. Overall, the MRSA-Screen test, GPS 106 card, and reference broth microdilution method had sensitivities of 95.7 (result at 10 min), 95.7, and 100%, respectively, and specificities of 92.8, 85.7, and 78.5%, respectively. Although the MRSA-Screen test required a slight procedural modification, both commercial methods achieved a sensitivity and specificity at detecting oxacillin resistance in CoNS at a level that was acceptable for clinical laboratory use.

PubMed Disclaimer

References

    1. Archer G L, Climo M W. Antimicrobial susceptibility of coagulase-negative staphylococci. Antimicrob Agents Chemother. 1994;38:2231–2237. - PMC - PubMed
    1. Chambers H F. Methicillin-resistant staphylococci. Clin Microbiol Rev. 1988;1:173–186. - PMC - PubMed
    1. Cormican M G, Wilke W W, Barrett M S, Pfaller M A, Jones R J. Phenotypic detection of mecA-positive staphylococcal blood stream isolates: high accuracy of simple disk diffusion tests. Diagn Microbiol Infect Dis. 1996;25:107–112. - PubMed
    1. Coudron P E, Jones D L, Dalton H P, Archer G L. Evaluation of laboratory tests for detection of methicillin-resistant Staphylococcus aureus and Staphylococcus epidermidis. J Clin Microbiol. 1986;24:764–769. - PMC - PubMed
    1. Geha D J, Uhl J R, Gustaferro C A, Persing D H. Multiplex PCR for identification of methicillin-resistant staphylococci in the clinical laboratory. J Clin Microbiol. 1994;32:1768–1772. - PMC - PubMed

Publication types

MeSH terms

LinkOut - more resources