A neuroprosthesis for high tetraplegia
Abstract
Background: This case report describes a neuroprosthesis that restored shoulder and elbow function in a 23-year-old man with chronic C3 complete tetraplegia. Before implementation of the neuroprosthesis, electrodiagnostic testing revealed denervation from C5 to T1, with the greatest degree of denervation in the C8 and T1 myotomes. Thirteen percutaneous intramuscular electrodes were implanted into muscles acting on the shoulder and elbow of one upper limb. Before functional testing, the subject underwent a conditioning regimen to maximize the strength and endurance of the implanted muscles.
Results: After completion of the 8-week exercise regimen, stimulated active range of motion against gravity included 60 degrees of shoulder abduction, 45 degrees of shoulder flexion, 10 degrees of shoulder external rotation with the shoulder passively abducted to 90 degrees, and 110 degrees of elbow flexion. Stimulated elbow extension lacked 20 degrees of full extension with gravity eliminated. After system setup, the subject was able to pick up mashed potatoes on a plate with a utensil and bring them to his mouth using the neuroprosthesis and a balanced forearm orthosis. A switch mounted on the headrest of the subject's wheelchair and a position sensor mounted on the contralateral shoulder allowed the subject to control movement of his upper limb.
Similar articles
-
Implanted neuroprosthesis for restoring arm and hand function in people with high level tetraplegia.Arch Phys Med Rehabil. 2014 Jun;95(6):1201-1211.e1. doi: 10.1016/j.apmr.2014.01.028. Epub 2014 Feb 20. Arch Phys Med Rehabil. 2014. PMID: 24561055 Free PMC article.
-
Spiral nerve cuff electrodes for an upper extremity neuroprosthesis.Conf Proc IEEE Eng Med Biol Soc. 2006;2006:3584-7. doi: 10.1109/IEMBS.2006.259625. Conf Proc IEEE Eng Med Biol Soc. 2006. PMID: 17947040
-
[The Freehand System: an implantable neuroprosthesis for functional electrostimulation of the upper extremity].Handchir Mikrochir Plast Chir. 2001 May;33(3):149-52. doi: 10.1055/s-2001-15129. Handchir Mikrochir Plast Chir. 2001. PMID: 11468891 German.
-
Development of an upper extremity FES system for individuals with C4 tetraplegia.IEEE Trans Rehabil Eng. 1996 Dec;4(4):264-70. doi: 10.1109/86.547926. IEEE Trans Rehabil Eng. 1996. PMID: 8973952 Review.
-
Functional electrical stimulation in tetraplegic patients to restore hand function.J Long Term Eff Med Implants. 2002;12(3):175-88. J Long Term Eff Med Implants. 2002. PMID: 12545943 Review.
Cited by
-
Use and evaluation of assistive technologies for upper limb function in tetraplegia.J Spinal Cord Med. 2022 Nov;45(6):809-820. doi: 10.1080/10790268.2021.1878342. Epub 2021 Feb 19. J Spinal Cord Med. 2022. PMID: 33606599 Free PMC article.
-
The science and engineering behind sensitized brain-controlled bionic hands.Physiol Rev. 2022 Apr 1;102(2):551-604. doi: 10.1152/physrev.00034.2020. Epub 2021 Sep 20. Physiol Rev. 2022. PMID: 34541898 Free PMC article. Review.
-
Implanted neuroprosthesis for restoring arm and hand function in people with high level tetraplegia.Arch Phys Med Rehabil. 2014 Jun;95(6):1201-1211.e1. doi: 10.1016/j.apmr.2014.01.028. Epub 2014 Feb 20. Arch Phys Med Rehabil. 2014. PMID: 24561055 Free PMC article.
-
Continuous neuronal ensemble control of simulated arm reaching by a human with tetraplegia.J Neural Eng. 2011 Jun;8(3):034003. doi: 10.1088/1741-2560/8/3/034003. Epub 2011 May 5. J Neural Eng. 2011. PMID: 21543840 Free PMC article.
-
An optimized proportional-derivative controller for the human upper extremity with gravity.J Biomech. 2015 Oct 15;48(13):3692-700. doi: 10.1016/j.jbiomech.2015.08.016. Epub 2015 Aug 29. J Biomech. 2015. PMID: 26358531 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous