Comparative Study
. 2001 Oct;29(10):1868-73.
doi: 10.1097/00003246-200110000-00003.

Mean cardiac output by thermodilution with a single controlled injection

Affiliations
Comparative Study

Mean cardiac output by thermodilution with a single controlled injection

J R Jansen et al. Crit Care Med. 2001 Oct.

Abstract

Objective: A new method to estimate mean cardiac output by thermodilution with a single duration-controlled injection was evaluated in patients.

Design: Prospective criterion standard study.

Setting: University hospital cardiac surgical intensive care unit and cardiac operation room.

Patients: Of 33 patients, 24 underwent coronary bypass graft surgery, four had a valve replacement, and five were treated in the intensive care unit.

Interventions: Interventions consisted of thermodilution cardiac output measurements. One single duration-controlled injection of cold fluid was used to calculate cardiac output. This controlled injection was performed with a duration equal to one whole ventilation cycle of the ventilator. An algorithm adapted to this duration-controlled injection calculated cardiac output. Moreover, this algorithm has properties to reduce errors caused by artificial ventilation and thermal noise.

Measurements and main results: In 33 patients, the averaged values of four measurements equally spread over the ventilatory cycle (phase-controlled) were compared with the values of two single duration-controlled measurements. The measurements were performed during periods of stable respiration and circulation. No significant difference was observed between the mean of four phase-controlled measurements and the mean of the two duration-controlled measurements. The cardiac output values in the intensive care patients were significantly higher compared with the two other patient groups (p <.05). The difference between the two methods could not be subdivided for the three patient groups (p >.05). The coefficient of variation of the single duration-controlled thermodilution measurements was significantly lower than the single phase-controlled measurements, 3% vs. 6% (p <.01).

Conclusions: One single duration-controlled injection thermodilution measurement is as accurate and repeatable as the mean of four phase-controlled measurements and is clinically feasible.