[Failure of acetabular cups fixed with cement and thick embedded wire mesh]
- PMID: 11685145
[Failure of acetabular cups fixed with cement and thick embedded wire mesh]
Abstract
Purpose of the study: We report our experience with the SKH(R) cemented acetabular cup, analyzing the clinical and radiographic consequences of using cement armed with a thick embedded wire mesh.
Material and methods: Between June 1990 and June 1991, 118 total hip arthroplasties (THAs) were performed in 110 patients (8 bilateral cases) using the SKH(R) cemented acetabular cup, a self-blocking cemented femoral stem with anterograde injection made of a Ti-6Al-7Nb alloy and a 28 mm ceramic polyethylene articulation. Mean follow-up for 97 implants (91 patients) was 7.1 years (3 - 9 years). Mean age at implantation was 65 years (36 - 85 years) (43 men and 48 women). THAs were performed for degenerative joint disease (n=79) including 67 cases of primary disease, necrosis (n=11), rheumatoid disease (n=6), and fracture of the femoral neck (n=1).
Results: For the 118 procedures, there were 4 complications: 2 dislocations (no revision), one failure of the ceramic head and one implant infection (with revision). As assessed by the Postel-Merle-d'Aubigné score as modified by Charnley for the mobility criterion, clinical outcome was excellent or good in 82% of the cases, fair in 11% and poor in 7%. According to the Massin criteria at last follow-up, there were 13 cup loosenings (13.4%, 10 migrations and 3 total lucent lines > 1 mm); revision procedures were performed in 6 cases. A lucent line postoperatively, observed in 51% of the cases (p=0.025) and involving all or part of zone 1 in 44% (p=0.047), was predictive of acetabular failure. Linear polyethylene wear as measured manually was 0.08 mm/year (average) for non-loosened cups and 0.2 mm/year for loosened cups; the difference was significant (p=0.001). On the contrary, the thickness of the cement, measured at the limits of the three De Lee and Charnley zones, cup tilt and size, age, weight and patient activity level had no effect on failure rate. Actuarial survival of the cup at 9 years was 86.3% taking aseptic loosening (with or without revision) as the failure criterion and 93.8% taking revision for aseptic loosening as the failure criterion.
Discussion: These results were rather disappointing. The thick layer of cement reinforced with the wire mesh produces a metal back effect with the known consequences in terms of loosening, cement-bone lucent line, and polyethylene wear. A modification is now undergoing thermal and mechanical assessment. A thinner and tighter mesh would appear to be preferable.
Similar articles
-
[Metal-on-metal hip replacement using Metasul cups cemented into Muller reinforcement rings after a mean 5-year (3-8) follow-up: improvement of acetabular fixation by comparing with direct cementation to bone].Rev Chir Orthop Reparatrice Appar Mot. 2008 Jun;94(4):346-53. doi: 10.1016/j.rco.2007.11.004. Epub 2008 Mar 4. Rev Chir Orthop Reparatrice Appar Mot. 2008. PMID: 18555860 French.
-
[Reconstruction of the acetabulum during replacement of the aseptically loosened polyethylene cup].Acta Chir Orthop Traumatol Cech. 2001;68(3):162-7. Acta Chir Orthop Traumatol Cech. 2001. PMID: 11706538 Czech.
-
[Results of the Evora dual mobility socket: five years follow-up].Rev Chir Orthop Reparatrice Appar Mot. 2008 Feb;94(1):37-42. doi: 10.1016/j.rco.2007.10.009. Epub 2008 Feb 1. Rev Chir Orthop Reparatrice Appar Mot. 2008. PMID: 18342028 French.
-
[Reliability of porous coating metal-backed cups: advantages and adverse effect of a posterior elevated rim polyethylene liner].Rev Chir Orthop Reparatrice Appar Mot. 2005 Sep;91(5):432-8. doi: 10.1016/s0035-1040(05)84360-0. Rev Chir Orthop Reparatrice Appar Mot. 2005. PMID: 16351000 Review. French.
-
Nine-year results of Müller cemented titanium Straight Stems in total hip replacement.Arch Orthop Trauma Surg. 2001 Jul;121(7):391-8. doi: 10.1007/s004020000254. Arch Orthop Trauma Surg. 2001. PMID: 11510904 Review.
Cited by
-
The augment-and-modular-cage revision system for reconstruction of severe acetabular defects-two-year clinical and radiographic results.Int Orthop. 2019 Oct;43(10):2269-2278. doi: 10.1007/s00264-018-4271-6. Epub 2018 Dec 11. Int Orthop. 2019. PMID: 30539219
-
Acetabular defect reconstruction in revision hip arthroplasty with a modular revision system and biological defect augmentation.Int Orthop. 2015 Apr;39(4):623-30. doi: 10.1007/s00264-014-2533-5. Epub 2014 Oct 4. Int Orthop. 2015. PMID: 25277762
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials