Evaluation of a near-infrared spectrometer (NIRO 300) for the detection of intracranial oxygenation changes in the adult head
- PMID: 11692006
- DOI: 10.1161/hs1101.098356
Evaluation of a near-infrared spectrometer (NIRO 300) for the detection of intracranial oxygenation changes in the adult head
Abstract
Background and purpose: The clinical application of intracranial near-infrared spectroscopy in adults has been hampered by concerns over contamination from extracranial tissues. The NIRO 300 (Hamamatsu Photonics) provides continuous online measurements of hemoglobin and cytochrome oxidase concentrations and a calculated tissue oxygen index (TOI). The present study seeks confirmation of the anatomic source of TOI in the adult cranium.
Methods: Sixty patients undergoing carotid endarterectomy were studied. The NIRO 300 was incorporated into an established multimodal monitoring system. TOI, oxyhemoglobin, and deoxyhemoglobin changes were assessed and compared with (1) frontal cutaneous laser-Doppler flowmetry and (2) transcranial Doppler measurement of the ipsilateral middle cerebral artery flow velocity.
Results: Changes in TOI were seen during cross-clamping of the carotid vessels in 49 patients (mean DeltaTOI=-9.4%, SD=7.1). Significant correlation was seen between TOI and flow velocity (r=0.56) but not with laser-Doppler flowmetry (r=0.13). In 31 patients, oxyhemoglobin and deoxyhemoglobin concentrations were recorded, showing significant changes during both external carotid artery and internal carotid artery clamping. A change in TOI was predominantly associated with internal carotid artery clamping (n=41). When TOI changed during external carotid artery clamping (n=8), significant blood pressure changes occurred, or extracranial-to-intracranial anastomosis was evident. In the absence of such variables, the sensitivity of TOI to intracranial and extracranial changes was 87.5% and 0%, respectively, and specificity was 100% and 0%, respectively.
Conclusions: The NIRO 300 reflects changes in cerebral tissue oxygenation when TOI is calculated, with a high degree of sensitivity and specificity.
Similar articles
-
Tissue oxygen index: thresholds for cerebral ischemia using near-infrared spectroscopy.Stroke. 2006 Nov;37(11):2720-5. doi: 10.1161/01.STR.0000244807.99073.ae. Epub 2006 Sep 28. Stroke. 2006. PMID: 17008623
-
Defining thresholds for critical ischemia by using near-infrared spectroscopy in the adult brain.J Neurosurg. 1998 Sep;89(3):389-94. doi: 10.3171/jns.1998.89.3.0389. J Neurosurg. 1998. PMID: 9724112
-
Measurement of cerebral tissue oxygenation in young healthy volunteers during acetazolamide provocation: a transcranial Doppler and near-infrared spectroscopy investigation.Adv Exp Med Biol. 2008;614:389-96. doi: 10.1007/978-0-387-74911-2_43. Adv Exp Med Biol. 2008. PMID: 18290350
-
Use of near-infrared spectroscopy in the adult.Philos Trans R Soc Lond B Biol Sci. 1997 Jun 29;352(1354):701-5. doi: 10.1098/rstb.1997.0052. Philos Trans R Soc Lond B Biol Sci. 1997. PMID: 9232858 Free PMC article. Review.
-
Perioperative uses of transcranial perfusion monitoring.Anesthesiol Clin. 2007 Sep;25(3):557-77, ix-x. doi: 10.1016/j.anclin.2007.05.002. Anesthesiol Clin. 2007. PMID: 17884708 Review.
Cited by
-
Cerebral Blood Deoxygenation by a Postural Change Detected by Near-Infrared Spectroscopy Has a Close Association with Cerebral Infarction.Brain Sci. 2022 Oct 21;12(10):1419. doi: 10.3390/brainsci12101419. Brain Sci. 2022. PMID: 36291352 Free PMC article.
-
Quantitative assessment of cytochrome C oxidase patterns in muscle tissue by the use of near-infrared spectroscopy (NIRS) in healthy volunteers.J Clin Monit Comput. 2022 Feb;36(1):271-278. doi: 10.1007/s10877-021-00648-6. Epub 2021 Jan 18. J Clin Monit Comput. 2022. PMID: 33459945 Clinical Trial.
-
Clinical Applications of Near-infrared Spectroscopy Monitoring in Cardiovascular Surgery.Anesthesiology. 2021 May 1;134(5):784-791. doi: 10.1097/ALN.0000000000003700. Anesthesiology. 2021. PMID: 33529323 Free PMC article. Review.
-
Head-up tilt and hyperventilation produce similar changes in cerebral oxygenation and blood volume: an observational comparison study using frequency-domain near-infrared spectroscopy.Can J Anaesth. 2012 Apr;59(4):357-65. doi: 10.1007/s12630-011-9662-8. Epub 2012 Jan 11. Can J Anaesth. 2012. PMID: 22234820 Free PMC article.
-
Multi-channel multi-distance broadband near-infrared spectroscopy system to measure the spatial response of cellular oxygen metabolism and tissue oxygenation.Biomed Opt Express. 2016 Oct 5;7(11):4424-4440. doi: 10.1364/BOE.7.004424. eCollection 2016 Nov 1. Biomed Opt Express. 2016. PMID: 27895985 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources