Transpedicular screw placement: image-guided versus lateral-view fluoroscopy: in vitro simulation
- PMID: 11698897
- DOI: 10.1097/00007632-200110010-00024
Transpedicular screw placement: image-guided versus lateral-view fluoroscopy: in vitro simulation
Abstract
Study design: In vitro evaluation of monitoring screw placement using an image-guided system compared with the routine use of an image intensifier.
Objectives: To compare a new computer-guided method of monitoring pedicle screw placement with a well-known method of monitoring using an image intensifier, using measurements of screw placement time and accuracy.
Summary of background data: Pedicle screw placement relies on the identification of anatomic landmarks for the point of entry of the pedicle. The direction and depth of screw penetration are guided by an intraoperative lateral-view image intensifier. The use of frameless stereotaxy for computer-aided pedicle screw placement may increase the accuracy and safety of the screw insertion. To the authors' knowledge, there are no published data comparing these systems on the basis of operative time and screw placement accuracy.
Method: Eight human cadaveric sections of five vertebrae each were used for an in vitro simulation of pedicle screw placement. Four spine surgeons were chosen to simulate the transpedicular screw placement. Each surgeon placed one screw into each pedicle of two spine sections (10 vertebrae, 20 screws). The surgeon was assisted by the lateral-view image intensifier on one spine section and by the navigational system on the second one. The four surgeons placed 80 pedicle screws. Forty screw placements were monitored by fluoroscopy and 40 by the image-guided navigational system. The time spent to place one screw was recorded, as well as the remarks by each surgeons on each method. Spines were rescanned, and the positions of the screws were compared between the group on which the image intensifier has been used and the group on which the navigational system had been used.
Results: In the image-guided technique group, one thoracic screw disrupted the lateral cortex of the pedicle, the average distance to the anterior wall of the body was 5 mm, and the average time for the insertion of one screw was 13.5 minutes. In the other group, two screws disrupted the inner cortex of a thoracic pedicle, the average distance to the anterior wall was 10.7 mm, and the average time for the insertion of one screw was 4 minutes.
Conclusions: In vitro computer-aided pedicle screw insertion is more accurate than lateral-view fluoroscopy in the thoracic spine. The main disadvantage is the time consumption compared with that required by lateral-view fluoroscopy. The total time of the surgical operation should be decreased with the future development of these techniques.
Similar articles
-
Pedicle screw placement in the thoracic spine: a comparison of image-guided and manual techniques in cadavers.Spine (Phila Pa 1976). 2005 Jun 15;30(12):E326-31. doi: 10.1097/01.brs.0000166621.98354.1d. Spine (Phila Pa 1976). 2005. PMID: 15959355
-
Image-guided spine surgery: a cadaver study comparing conventional open laminoforaminotomy and two image-guided techniques for pedicle screw placement in posterolateral fusion and nonfusion models.Spine (Phila Pa 1976). 2002 Nov 15;27(22):2503-8. doi: 10.1097/01.BRS.0000031274.34509.1E. Spine (Phila Pa 1976). 2002. PMID: 12435982
-
Accuracy of thoracic vertebral body screw placement using standard fluoroscopy, fluoroscopic image guidance, and computed tomographic image guidance: a cadaver study.Spine (Phila Pa 1976). 2003 Feb 15;28(4):402-13. doi: 10.1097/01.BRS.0000048461.51308.CD. Spine (Phila Pa 1976). 2003. PMID: 12590219
-
Computer navigation versus fluoroscopy-guided navigation for thoracic pedicle screw placement: a meta-analysis.Neurosurg Rev. 2016 Jul;39(3):385-91. doi: 10.1007/s10143-015-0679-2. Epub 2015 Dec 19. Neurosurg Rev. 2016. PMID: 26686852 Review.
-
Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature.Eur Spine J. 2008 Apr;17(4):564-75. doi: 10.1007/s00586-008-0584-2. Epub 2008 Jan 22. Eur Spine J. 2008. PMID: 18210169 Free PMC article. Review.
Cited by
-
Freehand Thoracic Pedicle Screw Placement: Review of Existing Strategies and a Step-by-Step Guide Using Uniform Landmarks for All Levels.Cureus. 2016 Feb 19;8(2):e501. doi: 10.7759/cureus.501. Cureus. 2016. PMID: 27014535 Free PMC article. Review.
-
Minimally invasive percutaneous transpedicular screw fixation: increased accuracy and reduced radiation exposure by means of a novel electromagnetic navigation system.Acta Neurochir (Wien). 2011 Mar;153(3):589-96. doi: 10.1007/s00701-010-0882-4. Epub 2010 Dec 14. Acta Neurochir (Wien). 2011. PMID: 21153669 Free PMC article.
-
Navigation-assisted fluoroscopy in minimally invasive direct lateral interbody fusion: a cadaveric study.SAS J. 2010 Dec 1;4(4):115-21. doi: 10.1016/j.esas.2010.09.002. eCollection 2010. SAS J. 2010. PMID: 25802659 Free PMC article.
-
High Speed, High Density Intraoperative 3D Optical Topographical Imaging with Efficient Registration to MRI and CT for Craniospinal Surgical Navigation.Sci Rep. 2018 Oct 5;8(1):14894. doi: 10.1038/s41598-018-32424-z. Sci Rep. 2018. PMID: 30291261 Free PMC article.
-
Usefulness of navigated O-arm(®) in a teaching center for spinal trauma.Asian J Neurosurg. 2016 Jul-Sep;11(3):298-302. doi: 10.4103/1793-5482.144179. Asian J Neurosurg. 2016. PMID: 27366260 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials