Improved hemodynamic effectiveness and associated clinical correlations of a new intermittent pneumatic compression system in patients with chronic venous insufficiency
- PMID: 11700495
- DOI: 10.1067/mva.2001.118822
Improved hemodynamic effectiveness and associated clinical correlations of a new intermittent pneumatic compression system in patients with chronic venous insufficiency
Abstract
Purpose: A new intermittent pneumatic compression device (SCD Response System) has recently been shown in healthy volunteers to have the ability to detect the postcompression refilling of the calf veins and to respond by initiating the subsequent cycle when these veins are full. This has proven to be more effective in expelling blood proximally than the conventional intermittent pneumatic compression device (SCD Sequel System). The aim of this study was to test the influence of venous disease on the postcompression refill time detected by means of the SCD Response and the effectiveness of the new system in expelling blood in patients who have venous reflux caused by post-thrombotic syndrome or varicose veins.
Methods: This open, controlled trial was conducted in an academic vascular unit with 10 patients who had post-thrombotic syndrome and 10 patients who had varicose veins. The new SCD Response System was tested against the existing SCD Sequel System in both legs in the supine, semirecumbent, and sitting positions. The refilling time sensed by means of the device was correlated with the venous filling index by using air plethysmography. The total volume of blood expelled per hour during compression was compared with that expelled by the SCD Sequel System in the same volunteers and in the same positions.
Results: An inverse association was found between the mean postcompression refilling time in the sitting position and the venous filling index of the apparently healthy or less severely affected leg (r = -0.52, P =.019), the refill time being significantly shorter in patients with advanced venous disease. The SCD Response System increased the volume expelled per hour in the post-thrombotic leg, when compared with the SCD Sequel System, by 109.9% (P =.005) in the supine position, by 85.1% (P =.009) in the semirecumbent position, and by 40.2% (P =.005) in the sitting position. The corresponding results in the more severely affected leg in patients with varicose veins were 71.9% (P =.005) in the supine position, 77.9% (P =.005) in the semirecumbent position, and 55.7% (P =.013) in the sitting position. Similar improved results were also found in the contralateral leg in both groups.
Conclusions: The deflation settings of the new SCD Response System are able to be adjusted selectively, correlating with the physiological severity of chronic venous insufficiency. By achieving more frequent compression cycles, the new system is more effective than the current one in expelling blood proximally, confirming our earlier findings in healthy volunteers. Further studies testing a possible improved efficacy in preventing deep venous thrombosis in this high-risk group are justified.
Similar articles
-
The efficacy of the new SCD response compression system in the prevention of venous stasis.J Vasc Surg. 2000 Nov;32(5):932-40. doi: 10.1067/mva.2000.110358. J Vasc Surg. 2000. PMID: 11054225 Clinical Trial.
-
The efficacy of a new portable sequential compression device (SCD Express) in preventing venous stasis.J Vasc Surg. 2005 Aug;42(2):296-303. doi: 10.1016/j.jvs.2005.03.031. J Vasc Surg. 2005. PMID: 16102630 Clinical Trial.
-
Can an intermittent pneumatic compression system monitor venous filling in the leg?J Med Eng Technol. 2008 May-Jun;32(3):221-7. doi: 10.1080/03091900701559424. J Med Eng Technol. 2008. PMID: 18432470
-
[Chronic venous insufficiency].Ned Tijdschr Geneeskd. 2002 Feb 2;146(5):199-203. Ned Tijdschr Geneeskd. 2002. PMID: 11851078 Review. Dutch.
-
Conception of the venous hemodynamics in the lower extremity.Angiology. 2006 Oct-Nov;57(5):556-63. doi: 10.1177/0003319706293117. Angiology. 2006. PMID: 17067977 Review.
Cited by
-
Compression therapy in venous diseases: physical assumptions and clinical effects.Postepy Dermatol Alergol. 2020 Dec;37(6):842-847. doi: 10.5114/ada.2019.86990. Epub 2019 Jul 26. Postepy Dermatol Alergol. 2020. PMID: 33603600 Free PMC article. Review.
-
Compression therapy for treating post-thrombotic syndrome.Cochrane Database Syst Rev. 2019 Sep 18;9(9):CD004177. doi: 10.1002/14651858.CD004177.pub2. Cochrane Database Syst Rev. 2019. PMID: 31531971 Free PMC article.
-
Comparison of clinical and physiological efficacies of different intermittent sequential pneumatic compression devices in preventing deep vein thrombosis: a prospective randomized study.Clin Orthop Surg. 2014 Dec;6(4):468-75. doi: 10.4055/cios.2014.6.4.468. Epub 2014 Nov 10. Clin Orthop Surg. 2014. PMID: 25436073 Free PMC article. Clinical Trial.
-
A prospective pilot study of thigh-administered intermittent pneumatic compression in the management of hard-to-heal lower limb venous and mixed aetiology ulcers.Int Wound J. 2019 Aug;16(4):940-945. doi: 10.1111/iwj.13125. Epub 2019 Apr 23. Int Wound J. 2019. PMID: 31016851 Free PMC article.
-
Compression therapy in venous diseases: current forms of compression materials and techniques.Postepy Dermatol Alergol. 2020 Dec;37(6):836-841. doi: 10.5114/ada.2019.86991. Epub 2019 Jul 26. Postepy Dermatol Alergol. 2020. PMID: 33603599 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources