Quantification of cerebral perfusion with "Real-Time" contrast-enhanced ultrasound
- PMID: 11714654
- DOI: 10.1161/hc4601.099400
Quantification of cerebral perfusion with "Real-Time" contrast-enhanced ultrasound
Abstract
Background: No noninvasive technique is currently capable of "real-time" assessment and monitoring of cerebral blood flow (CBF). We hypothesized that cerebral perfusion could be accurately measured and monitored in "real time" with contrast-enhanced ultrasound (CEU).
Methods and results: Cerebral perfusion was assessed in 9 dogs through a craniotomy with CEU at baseline and during hypercapnia and hypocapnia while normoxia was maintained. Cerebral microvascular blood volume (A), microbubble velocity (beta), and blood flow (Axbeta) were calculated from time-versus-acoustic intensity relations. Compared with baseline, hypercapnia and hypocapnia significantly increased and decreased CBF, respectively, as measured by CEU. These changes in blood flow were mediated by changes in both A and beta. A good correlation was found between Axbeta derived from CEU and CBF measured by radiolabeled microspheres (y=0.67x-0.04, r=0.91, P<0.001).
Conclusions: Changes in both cerebral microvascular blood volume and red blood cell velocity can be accurately assessed with CEU. Thus, CEU has the potential for bedside measurement and monitoring of cerebral perfusion in real time in patients with craniotomies or burr holes.
Similar articles
-
[Effects of acute cerebral ischemia on cerebral perfusion: quantitative evaluation by contrast-enhanced ultrasound in dogs].Nan Fang Yi Ke Da Xue Xue Bao. 2010 Jul;30(7):1518-21. Nan Fang Yi Ke Da Xue Xue Bao. 2010. PMID: 20650755 Chinese.
-
Cerebral perfusion in perinatal hypoxia and resuscitation assessed by transcranial contrast-enhanced ultrasound and 3 T MRI in newborn pigs.Invest Radiol. 2011 Nov;46(11):686-96. doi: 10.1097/RLI.0b013e3182266431. Invest Radiol. 2011. PMID: 21730873
-
Cerebral blood flow and cerebral blood flow velocity during angiotensin-induced arterial hypertension in dogs.Can J Anaesth. 1993 Aug;40(8):755-60. doi: 10.1007/BF03009772. Can J Anaesth. 1993. PMID: 8403159
-
Harmonic imaging--a new method for the sonographic assessment of cerebral perfusion.Eur J Ultrasound. 2001 Dec;14(2-3):103-13. doi: 10.1016/s0929-8266(01)00151-3. Eur J Ultrasound. 2001. PMID: 11704427 Review.
-
Contrast Enhanced Ultrasound Perfusion Imaging in Skeletal Muscle.J Cardiovasc Imaging. 2019 Jul;27(3):163-177. doi: 10.4250/jcvi.2019.27.e31. Epub 2019 May 20. J Cardiovasc Imaging. 2019. PMID: 31161755 Free PMC article. Review.
Cited by
-
Transcranial contrast-enhanced ultrasonography with Sonazoid in semiquantitative evaluation of brain perfusion.J Med Ultrason (2001). 2013 Apr;40(2):133-9. doi: 10.1007/s10396-013-0431-x. Epub 2013 Feb 28. J Med Ultrason (2001). 2013. PMID: 27277101
-
Contrast-enhanced ultrasonography with Sonazoid for evaluation of renal microcirculation.J Med Ultrason (2001). 2008 Dec;35(4):183-9. doi: 10.1007/s10396-008-0189-8. Epub 2008 Dec 16. J Med Ultrason (2001). 2008. PMID: 27278990
-
Cerebral blood flow, brain tissue oxygen, and metabolic effects of decompressive craniectomy.Neurocrit Care. 2012 Jun;16(3):478-84. doi: 10.1007/s12028-012-9685-1. Neurocrit Care. 2012. PMID: 22396191 Review.
-
Long-term high-fat diet induces hippocampal microvascular insulin resistance and cognitive dysfunction.Am J Physiol Endocrinol Metab. 2017 Feb 1;312(2):E89-E97. doi: 10.1152/ajpendo.00297.2016. Epub 2016 Nov 29. Am J Physiol Endocrinol Metab. 2017. PMID: 27899343 Free PMC article.
-
Contrast agent kinetics in the rabbit brain during exposure to therapeutic ultrasound.Ultrasound Med Biol. 2010 Jun;36(6):916-24. doi: 10.1016/j.ultrasmedbio.2010.03.005. Epub 2010 May 5. Ultrasound Med Biol. 2010. PMID: 20447757 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources