Evaluation of left ventricular endocardial volumes and ejection fractions computed from gated perfusion SPECT with magnetic resonance imaging: comparison of two methods
- PMID: 11725260
- DOI: 10.1067/mnc.2001.117173
Evaluation of left ventricular endocardial volumes and ejection fractions computed from gated perfusion SPECT with magnetic resonance imaging: comparison of two methods
Abstract
Background: Two methods of computing left ventricular volumes and ejection fraction (EF) from 8-frame gated perfusion single photon emission computed tomography (SPECT) were compared with each other and with magnetic resonance (MR) imaging.
Methods and results: Thirty-five subjects underwent 8-frame gated dual-isotope SPECT imaging and 12- to 16-frame gated MR imaging. Endocardial boundaries on short-axis MR images were hand traced by experts blinded to any SPECT results. Volumes and EF were computed with the use of Simpson's rule. SPECT images were analyzed for the same functional variables with the use of 2 automatic programs, Quantitative Gated SPECT (QGS) and the Emory Cardiac Toolbox (ECTb). The mean difference between MR and SPECT EF was 0.008 for ECTb and 0.08 for QGS. QGS showed a slight trend toward higher correlation for EF (r = 0.72, SE of the estimate = 0.08) than ECTb (r = 0.70, SE of the estimate = 0.09). For both SPECT methods, left ventricular volumes were similarly correlated with MR, although SPECT volumes were higher than MR values by approximately 30%.
Conclusions: QGS and ECTb values of cardiac function computed from 8-frame gated perfusion SPECT correlate very well with each other and correlate well with MR. Averaged over all subjects, ECTb measurements of EF are not significantly different from MR values but QGS significantly underestimates the MR values.
Similar articles
-
Quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPECT: MRI validation and comparison of the Emory Cardiac Tool Box with QGS and 4D-MSPECT.J Nucl Med. 2005 Aug;46(8):1256-63. J Nucl Med. 2005. PMID: 16085580 Clinical Trial.
-
Comparison of two software in gated myocardial perfusion single photon emission tomography, for the measurement of left ventricular volumes and ejection fraction, in patients with and without perfusion defects.Hell J Nucl Med. 2007 Jan-Apr;10(1):19-23. Hell J Nucl Med. 2007. PMID: 17450245
-
Evaluation of left ventricular ejection fraction by the quantitative algorithms QGS, ECTb, LMC and LVGTF using gated myocardial perfusion SPECT: investigation of relative accuracy.Nucl Med Commun. 2006 Apr;27(4):321-32. doi: 10.1097/01.mnm.0000202861.67293.95. Nucl Med Commun. 2006. PMID: 16531917
-
Are there normal values of phase analysis parameters for left ventricular dyssynchrony in patients with no structural cardiomyopathy?: a systematic review.Nucl Med Commun. 2019 Oct;40(10):980-985. doi: 10.1097/MNM.0000000000001068. Nucl Med Commun. 2019. PMID: 31469810
-
The expanding role of left ventricular functional assessment using gated myocardial perfusion SPECT: the supporting actor is stealing the scene.Eur J Nucl Med Mol Imaging. 2007 Jul;34(7):1107-22. doi: 10.1007/s00259-007-0405-6. Eur J Nucl Med Mol Imaging. 2007. PMID: 17384947 Review.
Cited by
-
Evaluation of the Quantitative Gated SPECT (QGS) software program in the presence of large perfusion defects.Int J Cardiovasc Imaging. 2005 Oct;21(5):519-29. doi: 10.1007/s10554-005-0274-7. Int J Cardiovasc Imaging. 2005. PMID: 16175442
-
Volume measurements in nuclear medicine gated SPECT and 4D echocardiography: validation using a dynamic cardiac phantom.Int J Cardiovasc Imaging. 2005 Apr-Jun;21(2-3):239-47; discussion 249-51. doi: 10.1007/s10554-004-4014-1. Int J Cardiovasc Imaging. 2005. PMID: 16015435
-
Impact of photon energy recovery on the assessment of left ventricular volume using myocardial perfusion SPECT.J Nucl Cardiol. 2004 May-Jun;11(3):312-7. doi: 10.1016/j.nuclcard.2004.03.001. J Nucl Cardiol. 2004. PMID: 15173778 Clinical Trial.
-
Assessment of left ventricular ejection fraction measured by quantitative gated SPECT: correlation with left ventriculography and first-pass radionuclide angiography.Int J Cardiovasc Imaging. 2006 Apr;22(2):223-30. doi: 10.1007/s10554-005-9014-2. Epub 2005 Oct 18. Int J Cardiovasc Imaging. 2006. PMID: 16231080
-
Noninvasive modalities for the assessment of left ventricular function: all are equal but some are more equal than others.J Nucl Cardiol. 2006 Jul;13(4):445-9. doi: 10.1016/j.nuclcard.2006.05.011. J Nucl Cardiol. 2006. PMID: 16919567 No abstract available.
References
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical