3D steady-state MR cisternography in CSF rhinorrhoea
Abstract
Purpose: To determine the utility of 3D steady-state MR cisternography in the demonstration and localisation of cerebrospinal fluid (CSF) leak in patients with clinically suspected CSF rhinorrhoea.
Material and methods: Six consecutive patients with clinically suspected CSF rhinorrhoea were examined with routine MR evaluation and MR cisternography (MRC). All MR examinations included fast spin-echo (SE) T1WI in axial and sagittal planes, fast SE T2WI in axial and coronal planes and fluid attenuated inversion recovery (FLAIR) images in the axial plane. 3D evaluation was done using the CISS technique with 0.7-mm thickness in the sagittal and coronal planes. The site and extent of the defect, and any brain herniation detected on MRC were correlated with surgical findings.
Results: In the 6 patients who underwent surgical exploration and repair, intraoperative findings correlated with the defect revealed by MRC in all cases.
Conclusion: In clinically suspected CSF rhinorrhoea, MRC is highly accurate in localising the site and extent of CSF fistula and may be used as the first investigation due to its efficacy and non-invasive nature.
Similar articles
-
The contribution of 3D-CISS and contrast-enhanced MR cisternography in detecting cerebrospinal fluid leak in patients with rhinorrhoea.Br J Radiol. 2010 Mar;83(987):225-32. doi: 10.1259/bjr/56838652. Epub 2009 Sep 1. Br J Radiol. 2010. PMID: 19723768 Free PMC article.
-
Diagnosis and Localization of Cerebrospinal Fluid Rhinorrhea: A Systematic Review.Am J Rhinol Allergy. 2022 May;36(3):397-406. doi: 10.1177/19458924211060918. Epub 2021 Nov 30. Am J Rhinol Allergy. 2022. PMID: 34846218 Free PMC article.
-
Non-Invasive and Minimally Invasive Imaging Evaluation of CSF Rhinorrhoea - a Retrospective Study with Review of Literature.Pol J Radiol. 2016 Feb 29;81:80-5. doi: 10.12659/PJR.895698. eCollection 2016. Pol J Radiol. 2016. PMID: 26985244 Free PMC article.
-
Evaluation of high-resolution CT and MR cisternography in the diagnosis of cerebrospinal fluid fistula.AJNR Am J Neuroradiol. 1998 Apr;19(4):633-9. AJNR Am J Neuroradiol. 1998. PMID: 9576647 Free PMC article.
-
Traumatic, iatrogenic, and spontaneous cerebrospinal fluid (CSF) leak: endoscopic repair.B-ENT. 2011;7 Suppl 17:47-60. B-ENT. 2011. PMID: 22338375 Review.
Cited by
-
MR ventriculocisternography by using 3D balanced steady-state free precession imaging: technical note.AJNR Am J Neuroradiol. 2005 May;26(5):1170-3. AJNR Am J Neuroradiol. 2005. PMID: 15891179 Free PMC article.
-
The contribution of 3D-CISS and contrast-enhanced MR cisternography in detecting cerebrospinal fluid leak in patients with rhinorrhoea.Br J Radiol. 2010 Mar;83(987):225-32. doi: 10.1259/bjr/56838652. Epub 2009 Sep 1. Br J Radiol. 2010. PMID: 19723768 Free PMC article.
-
Diagnosis and Localization of Cerebrospinal Fluid Rhinorrhea: A Systematic Review.Am J Rhinol Allergy. 2022 May;36(3):397-406. doi: 10.1177/19458924211060918. Epub 2021 Nov 30. Am J Rhinol Allergy. 2022. PMID: 34846218 Free PMC article.
-
Dural attachment of intracranial meningiomas: evaluation with contrast-enhanced three-dimensional fast imaging with steady-state acquisition (FIESTA) at 3 T.Neuroradiology. 2011 Jun;53(6):413-23. doi: 10.1007/s00234-010-0751-0. Epub 2010 Jul 29. Neuroradiology. 2011. PMID: 20668844
-
Tumor consistency of pituitary macroadenomas: predictive analysis on the basis of imaging features with contrast-enhanced 3D FIESTA at 3T.AJNR Am J Neuroradiol. 2014 Feb;35(2):297-303. doi: 10.3174/ajnr.A3667. Epub 2013 Aug 8. AJNR Am J Neuroradiol. 2014. PMID: 23928139 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical