A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra-high molecular weight polyethylene
- PMID: 11740757
- DOI: 10.1054/arth.2001.27234
A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra-high molecular weight polyethylene
Abstract
Cross-linked polyethylenes are being introduced widely in acetabular cups in hip prostheses as a strategy to reduce the incidence of wear debris-induced osteolysis. It will be many years before substantial clinical data can be collected on the wear of these new materials. Silane cross-linked polyethylene (XLPE) was introduced into clinical practice in a limited series of acetabular cups in 1986 articulating against 22.225-mm alumina ceramic femoral heads and showed reduced wear rates compared with conventionally sterilized (gamma irradiation in air) ultra-high molecular weight polyethylene (UHMWPE). We compared the wear of XLPE manufactured in 1986 with the wear of UHMWPE manufactured in 1986 in nonirradiated and irradiated forms. In the nonirradiated forms, the wear of XLPE was 3 times less than UHWMPE when articulating against smooth counterfaces. The nonirradiated materials did not show signs of oxidation. In the irradiated forms, only UHMWPE showed high levels of oxidation, and this caused a substantial increase in wear. Antioxidants added to XLPE during processing gave resistance to oxidative degradation. When sliding against scratched counterfaces, the wear of UHMWPE increased by a factor of 2 to 3 times. Against the same scratched counterfaces, the wear of XLPE increased dramatically by 30 to 200 times. This difference may be attributed to the reduction in toughness of XLPE. Clinically, XLPE has been articulated against damage-resistant ceramic heads, and this probably has been an important factor in contributing to reduced wear. New cross-linked polyethylenes differ considerably from XLPE. This study indicates that it is prudent to examine the wear of new polyethylenes under a range of conditions that may occur in vivo.
Comment in
-
A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra-high molecular weight polyethylene.J Arthroplasty. 2002 Dec;17(8):1078; author reply 1078-9. doi: 10.1016/s0883-5403(02)70001-6. J Arthroplasty. 2002. PMID: 12478524 No abstract available.
Similar articles
-
Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene.Biomed Mater Eng. 2001;11(1):23-35. Biomed Mater Eng. 2001. PMID: 11281576
-
Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions.Biomaterials. 2005 Jun;26(16):3259-67. doi: 10.1016/j.biomaterials.2004.07.070. Biomaterials. 2005. PMID: 15603821
-
Oxidation in ultrahigh molecular weight polyethylene and cross-linked polyethylene acetabular cups tested against roughened femoral heads in a hip joint simulator.Biomacromolecules. 2006 Jun;7(6):1912-20. doi: 10.1021/bm060007u. Biomacromolecules. 2006. PMID: 16768414
-
Implant wear mechanisms--basic approach.Biomed Mater. 2008 Dec;3(4):042001. doi: 10.1088/1748-6041/3/4/042001. Epub 2008 Sep 25. Biomed Mater. 2008. PMID: 18824778 Review.
-
Conventional ultra-high molecular weight polyethylene: a gold standard of sorts.Instr Course Lect. 2005;54:183-7. Instr Course Lect. 2005. PMID: 15948445 Review.
Cited by
-
Wear Debris Characterization and Corresponding Biological Response: Artificial Hip and Knee Joints.Materials (Basel). 2014 Feb 10;7(2):980-1016. doi: 10.3390/ma7020980. Materials (Basel). 2014. PMID: 28788496 Free PMC article. Review.
-
The effects of peroxide content on the wear behavior, microstructure and mechanical properties of peroxide crosslinked ultra-high molecular weight polyethylene used in total hip replacement.J Mater Sci Mater Med. 2008 Jun;19(6):2427-35. doi: 10.1007/s10856-008-3368-7. Epub 2008 Jan 25. J Mater Sci Mater Med. 2008. PMID: 18219557
-
Arthroplasty in Femoral Head Osteonecrosis.Hip Pelvis. 2014 Jun;26(2):65-73. doi: 10.5371/hp.2014.26.2.65. Epub 2014 Jun 30. Hip Pelvis. 2014. PMID: 27536561 Free PMC article. Review.
-
Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study.Int Orthop. 2010 Dec;34(8):1129-35. doi: 10.1007/s00264-009-0899-6. Epub 2009 Nov 1. Int Orthop. 2010. PMID: 19882338 Free PMC article. Clinical Trial.
-
Scratching vulnerability of conventional vs highly cross-linked polyethylene liners because of large embedded third-body particles.J Arthroplasty. 2012 May;27(5):742-9. doi: 10.1016/j.arth.2011.10.002. Epub 2011 Nov 23. J Arthroplasty. 2012. PMID: 22115764 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources