A reappraisal of pacemaker timing cycles pertaining to automatic mode switching
- PMID: 11752910
- DOI: 10.1023/a:1013298012367
A reappraisal of pacemaker timing cycles pertaining to automatic mode switching
Abstract
Automatic mode switching algorithms of dual chamber pacemakers require fundamental changes in the operation of pacemaker timing cycles to optimize detection of supraventricular tachyarrhythmias. The timing cycles related to mode switching are basically independent of the algorithm design. Blanking periods (when the sensing amplifier is temporarily disabled) should be optimized to a relatively small fraction of the pacing cycle to enhance atrial sensing and prevent far-field sensing. This review explains the function of the timing cycles pertaining to mode switching and proposes simpler terminology to facilitate the understanding of pacemaker function and electrographic interpretation of complex recordings.
Similar articles
-
[Mode-switching algorithms: programming and usefulness].Herz. 2001 Feb;26(1):2-17. doi: 10.1007/pl00002003. Herz. 2001. PMID: 11258106 German.
-
Mode switching for atrial tachyarrhythmias.Am J Cardiol. 1999 Mar 11;83(5B):202D-210D. doi: 10.1016/s0002-9149(98)01024-8. Am J Cardiol. 1999. PMID: 10089866 Review.
-
Timing cycles of permanent pacemakers.Cardiol Clin. 1992 Nov;10(4):593-608. Cardiol Clin. 1992. PMID: 1423375 Review.
-
Effect of the atrial blanking time on the detection of atrial fibrillation in dual chamber pacing.Pacing Clin Electrophysiol. 2001 Apr;24(4 Pt 1):496-9. doi: 10.1046/j.1460-9592.2001.00496.x. Pacing Clin Electrophysiol. 2001. PMID: 11341088
-
Failure of automatic mode switching: recognition and management.Pacing Clin Electrophysiol. 1997 Feb;20(2 Pt 1):268-75. doi: 10.1111/j.1540-8159.1997.tb06170.x. Pacing Clin Electrophysiol. 1997. PMID: 9058863
Cited by
-
Cardiac implantable devices during exercise: Normal function and troubleshooting.J Arrhythm. 2021 Mar 22;37(3):660-668. doi: 10.1002/joa3.12529. eCollection 2021 Jun. J Arrhythm. 2021. PMID: 34141019 Free PMC article. Review.
-
Preventricular far-field sensing in the atrial channel of dual chamber pacemakers--an occasional cause of inappropriate mode switch.J Interv Card Electrophysiol. 2004 Jun;10(3):231-5. doi: 10.1023/B:JICE.0000026917.82516.eb. J Interv Card Electrophysiol. 2004. PMID: 15133360
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical