Hemodynamic changes during a single treatment with the molecular adsorbents recirculating system in patients with acute-on-chronic liver failure
- PMID: 11753905
- DOI: 10.1053/jlts.2001.29108
Hemodynamic changes during a single treatment with the molecular adsorbents recirculating system in patients with acute-on-chronic liver failure
Abstract
The aim of this pilot study is to evaluate the circulatory safety of treatment with the molecular adsorbents recirculating system (MARS) by determining the effect on systemic hemodynamics of a single MARS treatment in patients with acute-on-chronic liver failure (AOCLF). In eight patients admitted with AOCLF, a single 10-hour MARS treatment was performed. Systemic hemodynamic variables were determined before and during treatment. Bilirubin and urea were monitored as measures of protein-bound and water-soluble toxins. During MARS treatment, mean arterial pressure increased from 67 +/- 9 to 76 +/- 6 mm Hg (P < .05). Systemic vascular resistance index increased from 757 +/- 134 to 884 +/- 183 dyne x s/cm(5)/m(2) (P < .05), whereas cardiac index remained constant (5.9 +/- 0.7 v 6.0 +/- 1.1 L/min/m(2)). No episode of dialysis-induced hypotension was observed. Systemic oxygen consumption remained constant (92 +/- 30 v 93 +/- 11 mL/min/m(2)). Bilirubin levels decreased from 537 +/- 192 to 351 +/- 106 micromol/L (P < .05), and urea levels, from 19.1 +/- 13.9 to 6.7 +/- 5.1 mmol/L (P < .05). In conclusion, MARS treatment proved safe in critically ill patients with no attributing side effects.
Similar articles
-
Cerebral blood flow velocity increases during a single treatment with the molecular adsorbents recirculating system in patients with acute on chronic liver failure.Liver Transpl. 2001 Aug;7(8):709-12. doi: 10.1053/jlts.2001.26059. Liver Transpl. 2001. PMID: 11510016
-
Effect of the molecular adsorbent recirculating system and Prometheus devices on systemic haemodynamics and vasoactive agents in patients with acute-on-chronic alcoholic liver failure.Crit Care. 2006;10(4):R108. doi: 10.1186/cc4985. Crit Care. 2006. PMID: 16859530 Free PMC article. Clinical Trial.
-
Prometheus versus molecular adsorbents recirculating system: comparison of efficiency in two different liver detoxification devices.Artif Organs. 2006 Apr;30(4):276-84. doi: 10.1111/j.1525-1594.2006.00215.x. Artif Organs. 2006. PMID: 16643386
-
Extracorporeal detoxification using the molecular adsorbent recirculating system for critically ill patients with liver failure.J Am Soc Nephrol. 2001 Feb;12 Suppl 17:S75-82. J Am Soc Nephrol. 2001. PMID: 11251037 Review.
-
Molecular adsorbent recirculating system for acute and acute-on-chronic liver failure: a meta-analysis.Liver Transpl. 2004 Sep;10(9):1099-106. doi: 10.1002/lt.20139. Liver Transpl. 2004. PMID: 15349999 Review.
Cited by
-
[Acute liver failure. Current aspects of diagnosis and therapy].Anaesthesist. 2004 Jun;53(6):511-30. doi: 10.1007/s00101-004-0682-4. Anaesthesist. 2004. PMID: 15098097 Review. German.
-
Clinical application of bioartificial liver support systems.Ann Surg. 2004 Aug;240(2):216-30. doi: 10.1097/01.sla.0000132986.75257.19. Ann Surg. 2004. PMID: 15273544 Free PMC article. Review.
-
When to use renal replacement therapy and bioartificial support for renal failure in patients with cirrhosis.Clin Liver Dis (Hoboken). 2013 Jun 21;2(3):116-119. doi: 10.1002/cld.184. eCollection 2013 Jun. Clin Liver Dis (Hoboken). 2013. PMID: 30992840 Free PMC article. Review. No abstract available.
-
Application of the Molecular Adsorbent Recirculating System in Type 1 Hepatorenal Syndrome in the Course of Alcohol-Related Acute on Chronic Liver Failure.Med Sci Monit. 2020 Jun 30;26:e923805. doi: 10.12659/MSM.923805. Med Sci Monit. 2020. PMID: 32602472 Free PMC article.
-
Bench-to-bedside review: current evidence for extracorporeal albumin dialysis systems in liver failure.Crit Care. 2007;11(3):215. doi: 10.1186/cc5922. Crit Care. 2007. PMID: 17567927 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous