Quiet MRI with novel acoustic noise reduction
- PMID: 11755088
- DOI: 10.1007/BF02678588
Quiet MRI with novel acoustic noise reduction
Abstract
Fast scan techniques, which are used to reduce scanning times, have raised scanning noise levels in magnetic resonance imaging (MRI) systems, resulting in greater patient discomfort and stress. It is well known that this noise is caused by vibration of the gradient coil due to the Lorentz forces generated by the current in the gradient coil, which is placed in a static magnetic field. We have confirmed that MRI noise can be substantially reduced by sealing the gradient coil in a vacuum chamber to block airborne vibration propagation, by supporting the gradient coil independently to block solid vibration propagation and by decreasing the eddy currents induced in RF coils, the RF shield and the static-field-magnet cryostat. Based on these findings, we have developed a silent MRI system in which scanning noise is markedly reduced under a wide range of scanning conditions.
Similar articles
-
[Quiet Technology of MRI].Igaku Butsuri. 2016;36(2):110-112. doi: 10.11323/jjmp.36.2_110. Igaku Butsuri. 2016. PMID: 28428453 Japanese.
-
Acoustic noise reduction in the NexGen 7 T scanner.Magn Reson Med. 2024 Nov;92(5):2261-2270. doi: 10.1002/mrm.30211. Epub 2024 Jul 14. Magn Reson Med. 2024. PMID: 39004827
-
Active-passive gradient shielding for MRI acoustic noise reduction.Magn Reson Med. 2005 May;53(5):1013-7. doi: 10.1002/mrm.20472. Magn Reson Med. 2005. PMID: 15844137
-
Sodium MRI radiofrequency coils for body imaging.NMR Biomed. 2016 Feb;29(2):107-18. doi: 10.1002/nbm.3392. Epub 2015 Sep 29. NMR Biomed. 2016. PMID: 26417667 Review.
-
Low-field MRI: An MR physics perspective.J Magn Reson Imaging. 2019 Jun;49(6):1528-1542. doi: 10.1002/jmri.26637. Epub 2019 Jan 13. J Magn Reson Imaging. 2019. PMID: 30637943 Free PMC article. Review.
Cited by
-
Quiet T1-Weighted Pointwise Encoding Time Reduction with Radial Acquisition for Assessing Myelination in the Pediatric Brain.AJNR Am J Neuroradiol. 2016 Aug;37(8):1528-34. doi: 10.3174/ajnr.A4747. Epub 2016 Apr 7. AJNR Am J Neuroradiol. 2016. PMID: 27056422 Free PMC article.
-
Masking level differences--a diffusion tensor imaging and functional MRI study.PLoS One. 2014 Feb 18;9(2):e88466. doi: 10.1371/journal.pone.0088466. eCollection 2014. PLoS One. 2014. PMID: 24558392 Free PMC article.
-
Silent echo-planar imaging for auditory FMRI.MAGMA. 2008 Sep;21(5):317-25. doi: 10.1007/s10334-008-0132-4. Epub 2008 Aug 21. MAGMA. 2008. PMID: 18716815
-
Applications in Awake Animal Magnetic Resonance Imaging.Front Neurosci. 2022 Apr 5;16:854377. doi: 10.3389/fnins.2022.854377. eCollection 2022. Front Neurosci. 2022. PMID: 35450017 Free PMC article. Review.
-
FMRI scanner noise interaction with affective neural processes.PLoS One. 2013 Nov 18;8(11):e80564. doi: 10.1371/journal.pone.0080564. eCollection 2013. PLoS One. 2013. PMID: 24260420 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical