Effect of vessel size on creation of hepatic radiofrequency lesions in pigs: assessment of the "heat sink" effect
- PMID: 11756085
- DOI: 10.2214/ajr.178.1.1780047
Effect of vessel size on creation of hepatic radiofrequency lesions in pigs: assessment of the "heat sink" effect
Abstract
Objective: The effect of vessels and their size on radiofrequency lesion creation in the liver was evaluated with respect to potential for vascular injury and perfusion-mediated "heat sink" effect.
Subjects and methods: Radiofrequency lesions targeted to tissue adjacent to a variety of vessels were created in vivo in the liver of 10 Yorkshire pigs. Postablation contrast-enhanced CT and then histopathologic analysis of the vessels and lesions were performed after sacrifice of the pigs. Degree of vascular injury and viability of perivascular hepatocytes were recorded and tabulated according to vessel size for both CT and histologic data sets.
Results: At CT, 42 (95%) of 44 veins greater than 3 mm remained patent, and four (20%) of 20 veins less than 3 mm were occluded. Heat sink effect, indicated by invagination of enhancing tissue between vessel and radiofrequency lesion, was observed in 32 (73%) of 44 veins greater than 3 mm and in zero of 20 veins less than 3 mm. On histopathology, 111 (100%) of 111 vessels less than 3 mm showed at least partial vessel wall injury, characterized by endothelial cell necrosis and luminal thrombus. In 24 vessels greater than 3 mm, the extent of vessel wall injury decreased with increasing vessel diameter. Viable perivascular tissue indicative of heat sink effect was identified in 12 of 24 veins greater than 3 mm, increasing to seven of seven veins greater than 5 mm. None of 96 vessels less than 2 mm and three of 111 vessels less than 3 mm showed any heat sink effect.
Conclusion: There appears to be a narrow transition zone for hepatic vessels at 2-4 mm, beyond which the heat sink effect was seen consistently and substantial vascular injury was rare.
Similar articles
-
Effect of vessel diameter on the creation of ovine lung radiofrequency lesions in vivo: preliminary results.J Surg Res. 2005 Mar;124(1):85-91. doi: 10.1016/j.jss.2004.09.008. J Surg Res. 2005. PMID: 15734484
-
The influence of radiofrequency ablation on hepatic vessels in porcine liver.Hepatogastroenterology. 2005 Mar-Apr;52(62):571-4. Hepatogastroenterology. 2005. PMID: 15816480
-
Microwave liver ablation: influence of hepatic vein size on heat-sink effect in a porcine model.J Vasc Interv Radiol. 2008 Jul;19(7):1087-92. doi: 10.1016/j.jvir.2008.03.023. Epub 2008 May 27. J Vasc Interv Radiol. 2008. PMID: 18589324
-
Imaging of pathology involving the space around the hepatic veins: "perivenous pattern".Diagn Interv Radiol. 2018 Mar-Apr;24(2):77-82. doi: 10.5152/dir.2018.17510. Diagn Interv Radiol. 2018. PMID: 29757146 Free PMC article. Review.
-
Non-neoplastic hepatic vascular diseases: spectrum of CT and MRI appearances.Clin Radiol. 2014 May;69(5):538-48. doi: 10.1016/j.crad.2013.12.017. Epub 2014 Feb 26. Clin Radiol. 2014. PMID: 24581966 Review.
Cited by
-
Characterization of the ablation zones produced by three commercially available systems from a single vendor for radiofrequency thermoablation in an ex vivo swine liver model.Vet Med Sci. 2020 Nov;6(4):1041-1048. doi: 10.1002/vms3.319. Epub 2020 Jul 1. Vet Med Sci. 2020. PMID: 32613738 Free PMC article.
-
Percutaneous ablation for perivascular hepatocellular carcinoma: Refining the current status based on emerging evidence and future perspectives.World J Gastroenterol. 2018 Dec 21;24(47):5331-5337. doi: 10.3748/wjg.v24.i47.5331. World J Gastroenterol. 2018. PMID: 30598578 Free PMC article. Review.
-
Experimental hepatic radiofrequency ablation using wet electrodes: electrode-to-vessel distance is a significant predictor for delayed portal vein thrombosis.Eur Radiol. 2006 Sep;16(9):1990-9. doi: 10.1007/s00330-006-0177-6. Epub 2006 Mar 16. Eur Radiol. 2006. PMID: 16541225
-
Local recurrence after hepatic radiofrequency coagulation: multivariate meta-analysis and review of contributing factors.Ann Surg. 2005 Aug;242(2):158-71. doi: 10.1097/01.sla.0000171032.99149.fe. Ann Surg. 2005. PMID: 16041205 Free PMC article. Review.
-
Conversion of the bronchial tree into a conforming electrode to ablate the lung nodule in a porcine model.Commun Med (Lond). 2023 Sep 29;3(1):129. doi: 10.1038/s43856-023-00362-1. Commun Med (Lond). 2023. PMID: 37775526 Free PMC article.