Abstract
Purpose: To assess and compare the reliability and reproducibility of retinal thickness measurements for the retinal thickness analyzer (RTA) and optical coherence tomography (OCT) in normal and edematous retina.
Methods: The authors measured the foveal thickness of 21 normal eyes and 9 eyes with macular edema with both methods in random order. With the RTA, the fovea was measured 10 times; with the OCT, six scans (one horizontal and five vertical cross-sections) of the fovea were obtained.
Results: Mean foveal thickness of normal eyes measured 153 microm with OCT and 181 microm with RTA (median for both methods 150 microm). Coefficients of variation (CV) within the same subjects were 10% (OCT) and 9% (RTA) reducing to 9% (OCT) and 7% (RTA) when scans were repeated only five times for both methods. The RTA, however, yielded an interpatient CV of 33% (OCT 17%), which was caused by several falsely high readings in normal individuals. In eyes with retinal thickening the OCT measured a mean of 324 microm with 15% intra- and 58% interpatient CV. The RTA yielded a mean of 403 microm with CV of 18% and 73%, respectively.
Conclusion: Both methods yield reproducible measurements of foveal thickness in normal individuals and individuals with macular edema. However, falsely high measurements may occur with the RTA, reducing its reliability as compared to the OCT.
Similar articles
-
Retinal thickness measurements with optical coherence tomography and the scanning retinal thickness analyzer.Retina. 2001;21(1):57-61. doi: 10.1097/00006982-200102000-00010. Retina. 2001. PMID: 11217931
-
Comparison between retinal thickness analyzer and optical coherence tomography for assessment of foveal thickness in eyes with macular disease.Am J Ophthalmol. 2002 Aug;134(2):240-51. doi: 10.1016/s0002-9394(02)01528-3. Am J Ophthalmol. 2002. PMID: 12140031
-
Retinal thickness in diabetic retinopathy: comparison of optical coherence tomography, the retinal thickness analyzer, and fundus photography.Retina. 2006 Jan;26(1):49-57. doi: 10.1097/00006982-200601000-00009. Retina. 2006. PMID: 16395139
-
Precision and reliability of retinal thickness measurements in foveal and extrafoveal areas of healthy and diabetic eyes.Invest Ophthalmol Vis Sci. 2008 Jun;49(6):2627-34. doi: 10.1167/iovs.07-0820. Invest Ophthalmol Vis Sci. 2008. PMID: 18515592
-
Retinal thickness study with optical coherence tomography in patients with diabetes.Invest Ophthalmol Vis Sci. 2002 May;43(5):1588-94. Invest Ophthalmol Vis Sci. 2002. PMID: 11980878
Cited by
-
Assessment of Macular Parameter Changes in Patients with Keratoconus Using Optical Coherence Tomography.J Ophthalmol. 2015;2015:245953. doi: 10.1155/2015/245953. Epub 2015 May 12. J Ophthalmol. 2015. PMID: 26064670 Free PMC article.
-
[Optical coherence tomography for macular edema. Classification, quantitative assessment, and rational usage in the clinical practice].Ophthalmologe. 2004 Aug;101(8):785-93. doi: 10.1007/s00347-004-1054-9. Ophthalmologe. 2004. PMID: 15459787 Review. German.
-
An Intelligent Optical Coherence Tomography-based System for Pathological Retinal Cases Identification and Urgent Referrals.Transl Vis Sci Technol. 2020 Aug 13;9(2):46. doi: 10.1167/tvst.9.2.46. eCollection 2020 Aug. Transl Vis Sci Technol. 2020. PMID: 32879756 Free PMC article.
-
Eccentricity and measurement variability and repeatability with the retinal thickness analyser.Br J Ophthalmol. 2004 Jan;88(1):62-5. doi: 10.1136/bjo.88.1.62. Br J Ophthalmol. 2004. PMID: 14693775 Free PMC article.
-
Imaging in glaucoma.Ophthalmol Clin North Am. 2004 Mar;17(1):33-52. doi: 10.1016/S0896-1549(03)00102-0. Ophthalmol Clin North Am. 2004. PMID: 15102512 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources