Prevention of pulmonary embolism by a foot sole pump
Abstract
We carried out a prospective, randomised study of 62 patients to determine the efficacy of a foot sole pump (the A-V Impulse System) for the prevention of pulmonary embolism (PE) after hip surgery. PE was assessed by pulmonary perfusion scintigraphy before and after operation. We defined a PE as any new scintigraphic defect which was larger than a bronchopulmonary segment. The incidence of PE was 55% in the control group and 21% in the treatment group. The foot sole pump significantly reduced the incidence of PE (p = 0.008) and we encountered no side-effects from its use.
Comment in
-
Prevention of pulmonary embolism by a foot sole pump.J Bone Joint Surg Br. 2002 Sep;84(7):1086-7; author reply 1087. J Bone Joint Surg Br. 2002. PMID: 12358380 No abstract available.