Evaluation of valve-related complications in patients with Sorin Bicarbon prosthesis: a seven-year experience
- PMID: 11767189
Evaluation of valve-related complications in patients with Sorin Bicarbon prosthesis: a seven-year experience
Abstract
Background and aim of the study: The study aim was to evaluate the medium-term results of aortic valve replacement (AVR), mitral valve replacement (MVR) and double valve replacement (DVR) with the Sorin Bicarbon prosthesis.
Methods: A total of 990 patients (568 men, 422 women; mean age 60+/-10 years; range: 20-86 years) was reviewed who received 1,108 Sorin Bicarbon prostheses between 1992 and 1998 at three institutions. AVR was performed in 541 patients (55%), MVR in 330 (33%) and DVR in 119 (12%). Concomitant procedures, mainly coronary artery grafting, were performed in 222 patients (22%). Follow up was 98% complete; total cumulative follow up was 3,091 patient-years.
Results: Hospital mortality was 3.7% (n = 37). There were 49 late deaths; actuarial survival at seven years was 88+/-2% after AVR, 86+/-5% after MVR, and 78+/-8% after DVR. At last follow up, 915 survivors were in NYHA functional class I or II. At seven years, actuarial freedom from valve-related deaths, valve thrombosis, embolism and bleeding respectively was 96+/-1%, 99+/-1%, 93+/-2% and 91+/-3% after AVR; 97+/-2%, 97+/-3%, 90+/-3% and 86+/-7% after MVR; and 92+/-6%, 98+/-1%, 64+/-2% and 82+/-2% after DVR. Reoperation was required in 20 patients (due to valve thrombosis in six, endocarditis in five and periprosthetic leak in nine). At seven years, actuarial freedom from reoperation was 97+/-1%, 96+/-2% and 84+/-9% after AVR, MVR and DVR, respectively; actuarial freedom from endocarditis was 99+/-1%, 99+/-1% and 95+/-5%. Nine patients experienced a nonstructural valve dysfunction (all periprosthetic leak), while no cases of structural failure were observed.
Conclusion: The Sorin Bicarbon prosthesis has shown good medium-term results with regard to clinical improvement, and low incidence of valve-related complications. Thus, it appears to be a reliable valve substitute when the use of a mechanical prosthesis is indicated.
Similar articles
-
Eighteen-year follow up after Hancock II bioprosthesis insertion.J Heart Valve Dis. 1999 Jan;8(1):16-24. J Heart Valve Dis. 1999. PMID: 10096477
-
Midterm evaluation of the Sorin Bicarbon heart valve prosthesis: single-center experience.Ann Thorac Surg. 2004 Feb;77(2):527-31. doi: 10.1016/S0003-4975(03)01347-X. Ann Thorac Surg. 2004. PMID: 14759432
-
Single center experience with the Sorin Bicarbon prosthesis: a 17-year clinical follow-up.J Thorac Cardiovasc Surg. 2014 Nov;148(5):2039-44. doi: 10.1016/j.jtcvs.2013.11.015. Epub 2013 Dec 9. J Thorac Cardiovasc Surg. 2014. PMID: 24332112
-
Clinical results of the ATS prosthetic valve in 240 implants and review of the literature.J Heart Valve Dis. 2001 Sep;10(5):628-35. J Heart Valve Dis. 2001. PMID: 11603603 Review.
-
Five-year evaluation of the Carpentier-Edwards porcine bioprosthesis.J Thorac Cardiovasc Surg. 1984 Sep;88(3):324-33. J Thorac Cardiovasc Surg. 1984. PMID: 6381888 Review.
Cited by
-
Comparative Long-Term Clinical Performance of Mechanical Aortic Valve Prostheses.JAMA Netw Open. 2024 Apr 1;7(4):e247525. doi: 10.1001/jamanetworkopen.2024.7525. JAMA Netw Open. 2024. PMID: 38639933 Free PMC article.
-
Fifteen-year experience with the Bicarbon heart valve prosthesis in a single center.J Cardiothorac Surg. 2015 Jun 28;10:89. doi: 10.1186/s13019-015-0294-x. J Cardiothorac Surg. 2015. PMID: 26123076 Free PMC article.
-
Clinical experience with the Bicarbon heart valve prosthesis.J Cardiothorac Surg. 2007 Jan 25;2:8. doi: 10.1186/1749-8090-2-8. J Cardiothorac Surg. 2007. PMID: 17254363 Free PMC article.