Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2001 Nov;15(9):925-8.
doi: 10.1089/089277901753284152.

Efficiency and efficacy of different intracorporeal ultrasonic lithotripsy units on a synthetic stone model

Affiliations
Comparative Study

Efficiency and efficacy of different intracorporeal ultrasonic lithotripsy units on a synthetic stone model

E N Liatsikos et al. J Endourol. 2001 Nov.

Abstract

Purpose: The efficiency and efficacy of the available intracorporeal ultrasonic lithotripters were compared in a stone model experiment.

Materials and methods: Plaster of Paris (POP) stone phantoms having ratios of 1:1, 1.5:1, and 2:1 with water were fabricated into cubes of various hardnesses weighing an average of 24.6 g. The stones were immersed in water in a plastic container, and continuous irrigation through a rigid nephroscope was used. Ultrasonic lithotripters from ACMI, Olympus, Storz, and Wolf manufacturers were evaluated for efficacy in breaking up the three POP concentrations. Time to complete stone fragmentation, occurrence of probe or tubing occlusion, and probe overheating were evaluated.

Results: Efficiency of fragmentation and time to fragmentation of the Storz lithotripter were significantly different from those of the Wolf (p = 0.01 and p = 0.02, respectively) and ACMI (p = 0.001 and p = 0.02, respectively) lithotripters. Comparison of the efficiency of fragmentation and time to fragmentation of the ACMI and Wolf lithotripters showed significant differences (p = 0.005 and p = 0.03, respectively) in favor of the Wolf device. The Olympus lithotriptor resulted in incomplete fragmentation of phantoms of all POP concentrations.

Conclusion: The Storz ultrasonic lithotriptor was found to have the lowest fragmentation time and highest efficiency in the fragmentation of phantom stones.