Tissue harmonic imaging improves organ visualization in trauma ultrasound when compared with standard ultrasound mode
- PMID: 11772669
- DOI: 10.1111/j.1553-2712.2002.tb01166.x
Tissue harmonic imaging improves organ visualization in trauma ultrasound when compared with standard ultrasound mode
Abstract
The focused abdominal sonography for trauma (FAST) examination is complicated by brightly lit trauma bays, limited time, and body habitus. Recently, new ultrasound (US) technology has become available that improves organ visualization in abdominal scans.
Objective: The hypothesis was that a new US mode, tissue harmonic (TH) imaging, improves visualization of critical organ relationships in the FAST examination by making use of previously unused frequencies. The authors performed a blind, prospective observational study to compare the images obtained in typical FAST views with those obtained in standard US and TH modes.
Methods: Blunt trauma patients presenting to a level I trauma center between April and September 2000 were enrolled on a convenience basis. Typical FAST views were obtained in standard and TH modes. The emergency ultrasonographer (EU) switched between modes for each view, optimizing the gain each time. Multiple digital still images were made with all indications of the mode used disguised. For each view on a patient, the best image in each mode was selected in a blinded fashion. Three experienced EUs, blinded to the mode used, rated each image pair for resolution, detail, and total image quality as previously defined on a ten-point Likert scale, 10 being the best for each category. Wilcoxon signed-ranks test, 95% confidence intervals (95% CIs), and interobserver correlation were calculated.
Results: A total of 76 image groups (39 of Morison's pouch, 20 splenorenal, and 17 bladder) from 52 patients were rated. Tissue harmonics produced improved resolution, detail, and quality when compared with the standard US mode, with median scores of 6.7 vs. 6.0, 6.7 vs. 6.0, and 6.3 vs. 6.0, respectively. The differences of 0.7 (95% CI = 0.4 to 0.93), 0.7 (95% CI = 0.4 to 0.93), and 0.33 (95% CI = 0.17 to 0.67) were statistically significant, with p = 0.0001, 0.0001, and 0.0003, respectively. There was good interobserver agreement (kappa = 0.74; 95% CI = 0.68 to 0.79).
Conclusions: Tissue harmonics produced FAST images higher in detail, resolution, and total image quality than standard-mode US images.
Similar articles
-
Ultrasound image quality comparison between an inexpensive handheld emergency department (ED) ultrasound machine and a large mobile ED ultrasound system.Acad Emerg Med. 2004 Jul;11(7):778-81. doi: 10.1197/j.aem.2003.12.030. Acad Emerg Med. 2004. PMID: 15231471 Clinical Trial.
-
Emergency department ultrasonography in the evaluation of hypotensive and normotensive children with blunt abdominal trauma.J Pediatr Surg. 2001 Jul;36(7):968-73. doi: 10.1053/jpsu.2001.24719. J Pediatr Surg. 2001. PMID: 11431759
-
Emergency physician use of ultrasonography in blunt abdominal trauma.Acad Emerg Med. 1996 Oct;3(10):931-7. doi: 10.1111/j.1553-2712.1996.tb03322.x. Acad Emerg Med. 1996. PMID: 8891039
-
Diagnostic accuracy of contrast enhanced ultrasound in patients with blunt abdominal trauma presenting to the emergency department: a systematic review and meta-analysis.Sci Rep. 2017 Jun 30;7(1):4446. doi: 10.1038/s41598-017-04779-2. Sci Rep. 2017. PMID: 28667280 Free PMC article.
-
Contrast-enhanced ultrasound for imaging blunt abdominal trauma - indications, description of the technique and imaging review.Ultraschall Med. 2012 Feb;33(1):60-7. doi: 10.1055/s-0031-1273442. Epub 2012 Jan 24. Ultraschall Med. 2012. PMID: 22274907 Review.
Cited by
-
Development and evaluation of a novel, real time mobile telesonography system in management of patients with abdominal trauma: study protocol.BMC Emerg Med. 2012 Dec 18;12:19. doi: 10.1186/1471-227X-12-19. BMC Emerg Med. 2012. PMID: 23249290 Free PMC article.
-
Diagnostic Accuracy of a Handheld Ultrasound vs a Cart-based Model: A Randomized Clinical Trial.West J Emerg Med. 2024 Mar;25(2):268-274. doi: 10.5811/westjem.17822. West J Emerg Med. 2024. PMID: 38596929 Free PMC article. Clinical Trial.
-
Prospective Comparison of Handheld Ultrasound Devices from Different Manufacturers with Respect to B-Scan Quality and Clinical Significance for Various Abdominal Sonography Questions.Diagnostics (Basel). 2023 Dec 8;13(24):3622. doi: 10.3390/diagnostics13243622. Diagnostics (Basel). 2023. PMID: 38132206 Free PMC article.
-
Comparing the image quality of tissue harmonic and conventional B-mode ultrasound of kidney in over-obese individuals.Electron Physician. 2018 Jul 25;10(7):7095-7100. doi: 10.19082/7095. eCollection 2018 Jul. Electron Physician. 2018. PMID: 30128101 Free PMC article.
-
Ultrasound Image Quality Comparison Between a Handheld Ultrasound Transducer and Mid-Range Ultrasound Machine.POCUS J. 2022 Apr 21;7(1):154-159. doi: 10.24908/pocus.v7i1.15052. eCollection 2022. POCUS J. 2022. PMID: 36896280 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous