Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2001;3(4):317-24.
doi: 10.1081/jcmr-100108585.

Prosthetic heart valves and annuloplasty rings: assessment of magnetic field interactions, heating, and artifacts at 1.5 Tesla

Affiliations

Prosthetic heart valves and annuloplasty rings: assessment of magnetic field interactions, heating, and artifacts at 1.5 Tesla

F G Shellock. J Cardiovasc Magn Reson. 2001.

Abstract

The purpose of this study was to determine the magnetic resonance (MR) safety aspects and artifacts for three different heart valve prostheses and two different annuloplasty rings that have not been evaluated previously in association with the 1.5-T MR environment. Ex vivo testing was performed using previously described techniques for the evaluation of magnetic field interactions (deflection angle and torque), heating (gel-filled phantom and fluoroptic thermometry; 15 min of MR imaging at a whole body-averaged specific absorption rate of 1.2 W/kg), and artifacts (using T1-weighted, spin echo, and gradient echo pulse sequences). One heart valve prosthesis and one annuloplasty ring showed no magnetic field interactions. Two heart valve prostheses and one annuloplasty ring displayed relatively minor magnetic field interactions (i.e., deflection angle < or = 6 degrees, torque, +1). Heating was < or = 0.7 degrees C for the five different implants. Artifacts varied depending on the amount and type of metal used to make the implants. For the three heart valve prostheses and two annuloplasty rings, the lack of substantial magnetic field interactions and relatively minor hearing indicated that MR procedures may be conducted safetly in patients with these implants using MR systems operating with static magneticfields of 1.5 T or less. Notably, these findings essentially apply to 54 different heart valve prostheses and 37 different annuloplasty rings (i.e., based on the various models and sizes available for these implants).

PubMed Disclaimer

Publication types

LinkOut - more resources