. 2002 Jan 8;58(1):97-103.
doi: 10.1212/wnl.58.1.97.

Complications of invasive video-EEG monitoring with subdural grid electrodes

Affiliations

Complications of invasive video-EEG monitoring with subdural grid electrodes

H M Hamer et al. Neurology. .

Abstract

Objective: To evaluate the risk factors, type, and frequency of complications during video-EEG monitoring with subdural grid electrodes.

Methods: The authors retrospectively reviewed the records of all patients who underwent invasive monitoring with subdural grid electrodes (n = 198 monitoring sessions on 187 patients; median age: 24 years; range: 1 to 50 years) at the Cleveland Clinic Foundation from 1980 to 1997.

Results: From 1980 to 1997, the complication rate decreased (p = 0.003). In the last 5 years, 19/99 patients (19%) had complications, including two patients (2%) with permanent sequelae. In the last 3 years, the complication rate was 13.5% (n = 5/37) without permanent deficits. Overall, complications occurred during 52 monitoring sessions (26.3%): infection (n = 24; 12.1%), transient neurologic deficit (n = 22; 11.1%), epidural hematoma (n = 5; 2.5%), increased intracranial pressure (n = 5; 2.5%), and infarction (n = 3; 1.5%). One patient (0.5%) died during grid insertion. Complication occurrence was associated with greater number of grids/electrodes (p = 0.021/p = 0.052; especially >60 electrodes), longer duration of monitoring (p = 0.004; especially >10 days), older age of the patient (p = 0.005), left-sided grid insertion (p = 0.01), and burr holes in addition to the craniotomy (p = 0.022). No association with complications was found for number of seizures, IQ, anticonvulsants, or grid localization.

Conclusions: Invasive monitoring with grid electrodes was associated with significant complications. Most of them were transient. Increased complication rates were related to left-sided grid insertion and longer monitoring with a greater number of electrodes (especially more than 60 electrodes). Improvements in grid technology, surgical technique, and postoperative care resulted in significant reductions in the complication rate.

PubMed Disclaimer

MeSH terms