Review of techniques to remove a Foley catheter when the balloon does not deflate
- PMID: 11796297
- DOI: 10.1016/s0090-4295(01)01483-2
Review of techniques to remove a Foley catheter when the balloon does not deflate
Abstract
Failure of a Foley balloon to deflate may be caused by a faulty valve mechanism, blockage of the inflation channel, or, rarely, crystallization of the fluid within the balloon. We discuss the approach to successful management and present a stepwise algorithm to remove the catheter safely and expeditiously.
Comment in
-
Foley catheter removal techniques.Urology. 2003 Jan;61(1):257; author reply 257. doi: 10.1016/s0090-4295(02)02152-0. Urology. 2003. PMID: 12559319 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources