Influence of the number of basis images and projection array on caries detection using tuned aperture computed tomography (TACT)
- PMID: 11803385
- DOI: 10.1038/sj/dmfr/4600656
Influence of the number of basis images and projection array on caries detection using tuned aperture computed tomography (TACT)
Abstract
Objectives: to determine if the number of basis images and spatial distribution of the projection array used for TACT slice generation influence observer performance in caries detection.
Methods: In the first experiment, 2, 4, 8 and 12 basis projections of each of 40 teeth were acquired using a CMOS digital radiography sensor. Projections were distributed radially in space using a 20 degree angular disparity. TACT slices were generated from the four subgroups of images, presented to eight observers, and viewed on a high-resolution monitor. Observers scored the presence/absence of caries using a 5-point confidence scale. Gold standard was histological examination of tooth sections. ROC curves measured observer diagnostic performance. ANOVA tested for significant differences between observers and experimental conditions. In the second experiment, the number of basis projections judged to be satisfactory for TACT slice generation was used. Horizontal and vertical linear arrays of projections were compared to the circular projection array.
Results: There was a statistically significant difference between the numbers of basis projections in the detection of both occlusal (P=0.006) and proximal caries (P=0.005). No significant difference was found between projection arrays in the detection of either occlusal (P=0.065) or proximal (P=0.515) caries.
Conclusions: The number of TACT basis projections significantly influences caries detection. Eight or more images should be used. Either linear-vertical, linear-horizontal or circular arrays of basis projections may be used for TACT slice generation in caries detection tasks.
Similar articles
-
Effect of angular disparity of basis images and projection geometry on caries detection using tuned-aperture computed tomography.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 Sep;92(3):353-60. doi: 10.1067/moe.2001.117812. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001. PMID: 11552158
-
Two- and three-dimensional imaging modalities for the detection of caries. A comparison between film, digital radiography and tuned aperture computed tomography (TACT).Dentomaxillofac Radiol. 1999 May;28(3):152-7. doi: 10.1038/sj/dmfr/4600430. Dentomaxillofac Radiol. 1999. PMID: 10740469
-
Generation of TACT image slices using different reconstruction algorithms: effects on natural caries detection.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001 Nov;92(5):576-81. doi: 10.1067/moe.2001.119146. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2001. PMID: 11709697
-
Diagnostic accuracy of Tuned Aperture Computed Tomography (TACT).Swed Dent J Suppl. 2003;(159):1-93. Swed Dent J Suppl. 2003. PMID: 12856440 Review.
-
Clinical validity of the interactive and low-dose three-dimensional dento-alveolar imaging system, tuned-aperture computed tomography.Bull Tokyo Dent Coll. 2003 Aug;44(3):159-67. doi: 10.2209/tdcpublication.44.159. Bull Tokyo Dent Coll. 2003. PMID: 14694831 Review.
Cited by
-
Radiographic modalities for diagnosis of caries in a historical perspective: from film to machine-intelligence supported systems.Dentomaxillofac Radiol. 2021 Jul 1;50(5):20210010. doi: 10.1259/dmfr.20210010. Epub 2021 Mar 4. Dentomaxillofac Radiol. 2021. PMID: 33661697 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical