Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator
Abstract
In this work, computational methods were utilized to optimize the field produced by the transverse electromagnetic (TEM) resonator in the presence of the human head at 8 Tesla. Optimization was achieved through the use of the classical finite difference time domain (FDTD) method and a TEM resonator loaded with an anatomically detailed human head model with a resolution of 2 mm x 2 mm x 2 mm. The head model was developed from 3D MR images. To account for the electromagnetic interactions between the coil and the tissue, the coil and the head were treated as a single system at all the steps of the model including, numerical tuning and excitation. In addition to 2, 3, 4, 6, and 10-port excitations, an antenna array concept was utilized by driving all the possible ports (24) of a 24-strut TEM resonator. The results show that significant improvement in the circularly polarized component of the transverse magnetic field could be obtained when using multiple ports and variable phase and fixed magnitude, or variable phase and variable magnitude excitations.
Similar articles
-
Analysis of B1 field profiles and SAR values for multi-strut transverse electromagnetic RF coils in high field MRI applications.Phys Med Biol. 2001 Oct;46(10):2545-55. doi: 10.1088/0031-9155/46/10/303. Phys Med Biol. 2001. PMID: 11686274
-
Electromagnetic perspective on the operation of RF coils at 1.5-11.7 Tesla.Magn Reson Med. 2005 Sep;54(3):683-90. doi: 10.1002/mrm.20596. Magn Reson Med. 2005. PMID: 16088934
-
Application of finite difference time domain method for the design of birdcage RF head coils using multi-port excitations.Magn Reson Imaging. 2000 Jul;18(6):733-42. doi: 10.1016/s0730-725x(00)00143-0. Magn Reson Imaging. 2000. PMID: 10930783
-
In-depth study of the electromagnetics of ultrahigh-field MRI.NMR Biomed. 2007 Feb;20(1):58-68. doi: 10.1002/nbm.1094. NMR Biomed. 2007. PMID: 17006885
-
A vision of 14 T MR for fundamental and clinical science.MAGMA. 2023 Apr;36(2):211-225. doi: 10.1007/s10334-023-01081-3. Epub 2023 Apr 10. MAGMA. 2023. PMID: 37036574 Free PMC article. Review.
Cited by
-
Imaging at ultrahigh magnetic fields: History, challenges, and solutions.Neuroimage. 2018 Mar;168:7-32. doi: 10.1016/j.neuroimage.2017.07.007. Epub 2017 Jul 8. Neuroimage. 2018. PMID: 28698108 Free PMC article. Review.
-
Manipulating transmit and receive sensitivities of radiofrequency surface coils using shielded and unshielded high-permittivity materials.MAGMA. 2018 Jun;31(3):355-366. doi: 10.1007/s10334-017-0657-5. Epub 2017 Nov 6. MAGMA. 2018. PMID: 29110240 Free PMC article.
-
Approaching ultimate intrinsic specific absorption rate in radiofrequency shimming using high-permittivity materials at 7 Tesla.Magn Reson Med. 2018 Jul;80(1):391-399. doi: 10.1002/mrm.27022. Epub 2017 Nov 28. Magn Reson Med. 2018. PMID: 29193307 Free PMC article.
-
Perivascular Space Imaging at Ultrahigh Field MR Imaging.Magn Reson Imaging Clin N Am. 2021 Feb;29(1):67-75. doi: 10.1016/j.mric.2020.09.005. Epub 2020 Nov 2. Magn Reson Imaging Clin N Am. 2021. PMID: 33237016 Free PMC article. Review.
-
Array-optimized composite pulse for excellent whole-brain homogeneity in high-field MRI.Magn Reson Med. 2007 Mar;57(3):470-4. doi: 10.1002/mrm.21172. Magn Reson Med. 2007. PMID: 17326169 Free PMC article.
Publication types