Validity and reliability of a new portable telemetric calorimeter designed to measure oxygen consumption and carbon dioxide production
- PMID: 11806467
Validity and reliability of a new portable telemetric calorimeter designed to measure oxygen consumption and carbon dioxide production
Abstract
Objective: The purpose of the present study was to evaluate the precision and the accuracy of a new portable telemetric calorimeter developed for the measurement of oxygen consumption, carbon dioxide production and pulmonary ventilation.
Design: An in vitro validation study.
Methods: An experimental protocol was designed to generate a series of tidal volumes (from 100 ml to 6000 ml at 200 ml steps) at different respiratory rates (5, 8, 14, 20, 40, 60, 75 breaths-min(-1)). For this purpose, a standardized pulmonary waveform generator system was utilized. Moreover, in order to evaluate the measure of O2 and CO2 concentrations at different temperatures (-10 degrees, 0 degrees, 15 degrees, 25 degrees, 40 degrees C), the instrument was placed inside an adjustable temperature and humidity chamber.
Results: The accuracy of flow measurements was within +/-3% in respect to the real values throughout the entire range of physiological values, whereas the measurements of gas concentrations were within +/-1% between 0 and 25 degrees C.
Conclusions: The new portable telemetric calorimeter represents an easy to use instrument for the measure of energy expenditure during the activities of daily living.
Similar articles
-
An in vitro evaluation of an instrument designed to measure oxygen consumption and carbon dioxide production during mechanical ventilation.Crit Care Med. 1994 Dec;22(12):1995-200. Crit Care Med. 1994. PMID: 7988139
-
Validation of the VO2000 calorimeter for measuring resting metabolic rate.Clin Nutr. 2006 Aug;25(4):687-92. doi: 10.1016/j.clnu.2006.01.002. Epub 2006 May 15. Clin Nutr. 2006. PMID: 16698140
-
A whole body transportable indirect calorimeter for human use in the tropics.Eur J Clin Nutr. 1990 Oct;44(10):725-31. Eur J Clin Nutr. 1990. PMID: 2269251
-
Some mathematical and technical issues in the measurement and interpretation of open-circuit indirect calorimetry in small animals.Int J Obes (Lond). 2006 Sep;30(9):1322-31. doi: 10.1038/sj.ijo.0803280. Epub 2006 Jun 27. Int J Obes (Lond). 2006. PMID: 16801931 Review.
-
Metabolic measurements in the critically ill.Crit Care Clin. 1995 Jan;11(1):169-97. Crit Care Clin. 1995. PMID: 7736266 Review.
Cited by
-
Physical activity energy expenditure measured using a portable telemetric device in comparison with a mass spectrometer.Br J Sports Med. 2003;37(5):445-7. doi: 10.1136/bjsm.37.5.445. Br J Sports Med. 2003. PMID: 14514539 Free PMC article.