. 2002 Jan;127(1):144-8.
doi: 10.1039/b108240j.

Quantitative determination of acetylsalicylic acid and acetaminophen in tablets by FT-Raman spectroscopy

Affiliations

Quantitative determination of acetylsalicylic acid and acetaminophen in tablets by FT-Raman spectroscopy

Roman Szostak et al. Analyst. 2002 Jan.

Abstract

A procedure for quantitative determination of acetylsalicylic acid and acetaminophen in pharmaceuticals by PLS (partial least squares) and PCR (principal component regression) treatment of FT (Fourier transform)-Raman spectroscopic data is proposed. The proposed method was tested on powdered samples. Three chemometric models were built: the first, for samples consisting of an active substance diluted by lactose, starch and talc; the second, in which a simple inorganic salt was applied as an internal standard and additions were not taken into account; and the third, in which a model was constructed for a commercial pharmaceutical, where all constituents of the tablet were known. By utilising selected spectral ranges and by changing the chemometric conditions it is possible to carry out fast and precise analysis of the active component content in medicines on the basis of the simplified chemometric models. The proposed method was tested on five commercial tablets. The results were compared with data obtained by intensity ratio and pharmacopoeial methods. To appraise the quality of the models, the relative standard error of predictions (RSEPs) were calculated for calibration and prediction data sets. These were 0.7-2.0% and 0.8-2.3%, respectively, for the different PLS models. Application of these models to the Raman spectra of commercial tablets containing acetylsalicylic acid gave RSEP values of 1.3-2.0% and a mean accuracy of 1.2-1.7% with a standard deviation of 0.6-1.2%.

PubMed Disclaimer

MeSH terms

LinkOut - more resources