Oversizing of self-expanding stents: influence on the development of neointimal hyperplasia of the carotid artery in a canine model
- PMID: 11827884
- PMCID: PMC7975499
Oversizing of self-expanding stents: influence on the development of neointimal hyperplasia of the carotid artery in a canine model
Abstract
Background and purpose: In carotid artery stent placement, marked oversizing of the stent relative to the internal carotid artery lumen is common. This study was performed to determine the influence of using oversized self-expanding nitinol stents on neointimal hyperplasia.
Methods: In six greyhound dogs, 24 self-expanding nitinol stents (eight SMART stents, eight Easy Wallstents, eight Sinus-Flex stents) were inserted into both common carotid arteries (CCAs). In each CCA, two stents were deployed; a stent of the appropriate diameter was implanted distally and an oversized stent proximally. After 4 months, transverse sections of each stent were examined histologically and at computerized image analysis. Neointimal hyperplasia was determined as the proportion of the residual diameter of the patent vessel lumen compared with the stent lumen.
Results: The amount of neointimal hyperplasia did not differ between the normal-sized and oversized stents. The mean preserved luminal diameter (+/-SEM) with normal- and oversized stents, respectively, were as follows: Easy Wallstent, 94% +/- 1.0 and 96% +/- 1.5; SMART stent, 92% +/- 1.6 and 93% +/- 1.8; and Sinus-Flex stent, 93% +/- 2.7 and 93% +/- 2.6. The mean preserved patent lumen with the 12 normal-sized stents (93% +/-1.0) was not significantly different from that of the 12 oversized stents (94% +/- 1.1, P =.502).
Conclusion: Under experimental conditions, use of self-expanding stents oversized by 30-40% appeared to result in neointimal hyperplasia comparable to that caused by normal-sized stents. All three stent types appeared to have similarly low neointimal responses.
Figures
References
-
- Wholey MH, Wholey M, Mathias K, et al. Global experience in cervical carotid artery stent placement. Catheter Cardiovasc Interv 2000;50:160–167 - PubMed
-
- Yadav JS, Roubin GS, Iyer S, et al. Elective stenting of the extracranial carotid arteries. Circulation 1997;95:376–381 - PubMed
-
- Kirsch EC, Khangure MS, van Schie GP, Lawrence-Brown MM, Stewart-Wynne EG, McAuliffe W. Carotid artery stenting: results and follow-up in 53 patients. Radiology 2001;220:737–744 - PubMed
-
- Mathur A, Dorros G, Iyer SS, Vitek JJ, Yadav SS, Roubin GS. Palmaz stent compression in patients following carotid artery stenting. Cathet Cardiovasc Diagn 1997;41:137–140 - PubMed
-
- Johnson SP, Fujitani RM, Leyendecker JR, Joseph FB. Stent deformation and intimal hyperplasia complicating treatment of a post-carotid endarterectomy intimal flap with a Palmaz stent. J Vasc Surg 1997;25:764–768 - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical