Angiography-guided routine coronary stent implantation results in suboptimal dilatation
- PMID: 11863311
- DOI: 10.1177/000331970205300109
Angiography-guided routine coronary stent implantation results in suboptimal dilatation
Abstract
The resistance of the atherosclerotic lesion counteracts the expansion of the stent, resulting in suboptimal stent expansion. Intravascular ultrasound provides more precise information on stent expansion than coronary angiography but adds cost and time to the percutaneous transluminal coronary angiography procedure. The aim of this study was to evaluate the need for intravascular ultrasound at routine angiography-guided high pressure stent implantation by comparing stent expansion with predefined intracoronary ultrasound criteria for optimal stent implantation. In 32 patients, 48 stents (35 NIR, 12 AVE, and 1 Cordis) were implanted in A, B, and C stenoses using a high-pressure inflation technique until an optimal result was achieved according to angiography. Stent expansion was then evaluated by intravascular ultrasound as minimal lumen diameter, minimal lumen area, proximal and distal stent area, and a minimal lumen area symmetry index. These variables were then compared with the nominal stent size in vitro. Finally the stents were also evaluated with respect to the MUSIC criteria, ie, strict criteria regarding symmetry, apposition, and vessel geometry according to intravascular ultrasound after stent expansion. Forty-five stents could be completely analyzed. The mean balloon inflation pressure was 12.8 (range, 10-17) atm. The nominal stent size was not achieved in any patient. Minimal lumen diameter attained 77% and minimal lumen area 78% of expected nominal values (p<0.0001), distal stent area 88% (p < 0.001), and proximal stent area 92% (ns). Application of the MUSIC criteria showed that almost all stents (96%) had good stent apposition and symmetry index. Optimal proximal stent entrance was found in 70%. Optimal minimal lumen area in comparison to the reference areas was present in 41%. This lead to fulfilling of all MUSIC criteria in 47% of the stents. If nominal stent size had been achieved, symmetry index and apposition would have been fulfilled in all cases and optimal minimal lumen area increased to 75%. Acceptable proximal entrance however would have decreased to 55% and the fulfillment of all MUSIC criteria would increase only to 52%. In routine angiography-guided stent implantation in stenoses with a wide range of severities using modern stents and high pressure inflation technique to reach a visually optimal result, the nominal stent size was never achieved mainly due to residual intrastent stenosis. If nominal stent size had been achieved, the results would have improved only marginally and would still be suboptimal in almost half of the stents. These results highlight the shortcoming of angiography and the need for intravascular ultrasound in choosing correct stent size.
Similar articles
-
Standardized angiographically guided over-dilatation of stents using high pressure technique optimize results without increasing risks.J Invasive Cardiol. 2002 May;14(5):221-6. J Invasive Cardiol. 2002. PMID: 11983940 Clinical Trial.
-
Is quantitative angiography sufficient to guide stent implantation? A comparison with three-dimensional reconstruction of intracoronary ultrasound images.G Ital Cardiol. 1997 Apr;27(4):328-36. G Ital Cardiol. 1997. PMID: 9199951 Clinical Trial.
-
Additional luminal area gain by intravascular ultrasound guidance after coronary stent implantation with high inflation pressure.Int J Card Imaging. 1997 Aug;13(4):311-21. doi: 10.1023/a:1005703626872. Int J Card Imaging. 1997. PMID: 9306145 Clinical Trial.
-
[New methods of coronary imaging II. Intracoronary ultrasonography in clinical practice].Ital Heart J Suppl. 2001 Jun;2(6):579-92. Ital Heart J Suppl. 2001. PMID: 11460831 Review. Italian.
-
Intravascular Ultrasound for Guidance and Optimization of Percutaneous Coronary Intervention.Interv Cardiol Clin. 2018 Jul;7(3):315-328. doi: 10.1016/j.iccl.2018.03.002. Epub 2018 Jun 29. Interv Cardiol Clin. 2018. PMID: 29983144 Review.
Cited by
-
Expert Opinion: Optimising Stent Deployment in Contemporary Practice: The Role of Intracoronary Imaging and Non-compliant Balloons.Interv Cardiol. 2017 Sep;12(2):81-84. doi: 10.15420/icr.2017:12:1. Interv Cardiol. 2017. PMID: 29588734 Free PMC article.
-
Efficacy of non-compliant balloon post-dilation in optimization of contemporary stents: A digital stent enhancement study.Int J Cardiol Heart Vessel. 2014 Mar 19;3:43-48. doi: 10.1016/j.ijchv.2014.03.006. eCollection 2014 Jun. Int J Cardiol Heart Vessel. 2014. PMID: 29450169 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous