Quantification of left ventricular diastolic pressure-volume relations during routine cardiac catheterization by two-dimensional digital echo quantification and left ventricular micromanometer
- PMID: 11875385
- DOI: 10.1067/mje.2002.118174
Quantification of left ventricular diastolic pressure-volume relations during routine cardiac catheterization by two-dimensional digital echo quantification and left ventricular micromanometer
Abstract
Background: Currently there is no simple clinical method for quantifying the left ventricular (LV) diastolic pressure-volume relation. Echocardiographic-automated endocardial border detection, however, may be combined with LV micromanometer to construct LV pressure-volume loops. We investigated the feasibility of on-line display and sampling of LV pressure-volume loops by such an approach. For this purpose we used a new echocardiographic digital echo quantification (DEQ) method in combination with LV pressures on-line and in real-time.
Methods: Eighteen patients were screened by conventional echocardiography and DEQ. Ten of the patients with high quality images were included in the study. Left ventricular pressures and volumes were recorded simultaneously and were displayed on-line as pressure-volume loops. Changes in LV volume were induced by intravenous saline. Left ventricular chamber compliance was estimated as change in volume divided by change in pressure from minimum diastolic pressure to end-diastolic pressure (average LV chamber compliance).
Results: Left ventricular pressure-volume loops were displayed on-line during the examination. When compared with the Simpson's method, DEQ underestimated end-diastolic volume (EDV) by 35% and overestimated end-systolic volume (ESV) by 14%. Beat-to-beat variability for ESV and EDV were 7.4% +/- 0.8% and 7.2% +/- 0.7 %, respectively. Volume loading increased LV end-diastolic pressure (LVEDP) from 14.0 +/- 1.6 to 24.7 +/- 2.0 mm Hg (P <.05) and EDV from 79 +/- 10 to 85 +/- 11 mL (NS), and decreased LV chamber compliance from 4.0 +/- 0.7 to 2.0 +/- 0.3 mL/mm Hg (P <.05).
Conclusion: The current study demonstrates that LV pressure-volume loops can be displayed and evaluated in real-time during routine cardiac catheterization. This may represent a clinically useful method for identifying patients with reduced chamber compliance. The underestimation of the volumes by DEQ compared with the Simpson's method suggests that further refinements should be performed to improve the endocardial border detection algorithm.
Similar articles
-
Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography.Cardiovasc Ultrasound. 2018 Jan 23;16(1):2. doi: 10.1186/s12947-017-0121-8. Cardiovasc Ultrasound. 2018. PMID: 29357888 Free PMC article.
-
Estimation of left ventricular compliance using on-line echocardiographic automated border detection and pressure data.Int J Card Imaging. 1994 Jun;10(2):103-11. doi: 10.1007/BF01137705. Int J Card Imaging. 1994. PMID: 7963748
-
Estimation of left ventricular chamber and stroke volume by limited M-mode echocardiography and validation by two-dimensional and Doppler echocardiography.Am J Cardiol. 1996 Oct 1;78(7):801-7. doi: 10.1016/s0002-9149(96)00425-0. Am J Cardiol. 1996. PMID: 8857486 Clinical Trial.
-
Diastolic function in various forms of left ventricular hypertrophy: contribution of active Doppler stress echo.Int J Sports Med. 1996 Nov;17 Suppl 3:S184-90. doi: 10.1055/s-2007-972922. Int J Sports Med. 1996. PMID: 9119541 Review.
-
Echocardiographic assessment of left ventricular diastolic pressure.Curr Opin Crit Care. 2019 Jun;25(3):252-258. doi: 10.1097/MCC.0000000000000601. Curr Opin Crit Care. 2019. PMID: 30865612 Review.
Cited by
-
Closing the loop: modelling of heart failure progression from health to end-stage using a meta-analysis of left ventricular pressure-volume loops.PLoS One. 2014 Dec 5;9(12):e114153. doi: 10.1371/journal.pone.0114153. eCollection 2014. PLoS One. 2014. PMID: 25479594 Free PMC article.
-
Understanding the need of ventricular pressure for the estimation of diastolic biomarkers.Biomech Model Mechanobiol. 2014 Aug;13(4):747-57. doi: 10.1007/s10237-013-0531-y. Epub 2013 Oct 4. Biomech Model Mechanobiol. 2014. PMID: 24092256 Free PMC article.
-
Assessment of left ventricular volume and function by integration of simplified 3D echocardiography, tissue harmonic imaging and automated extraction of endocardial borders.Int J Cardiovasc Imaging. 2004 Jun;20(3):191-202. doi: 10.1023/b:caim.0000021948.96454.3a. Int J Cardiovasc Imaging. 2004. PMID: 15139533
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous