Abstract
Plasma human immunodeficiency virus RNA in 491 clinical specimens was measured by LCx. There was a strong correlation with the results provided by other methods (r(2) values of 0.93 with Cobas-Monitor version 1.5 and of 0.95 with Quantiplex version 3.0). However, values were uniformly higher with LCx than with Quantiplex when non-B subtypes were tested.
Figures
References
-
- Brodine, S. K., J. R. Mascola, P. J. Weiss, S. I. Ito, K. R. Porter, A. W. Artenstein, F. C. Garland, F. E. McCutchan, and D. S. Burke. 1995. Detection of diverse HIV-1 genetic subtypes in the USA. Lancet 346:1198-1999. - PubMed
-
- Collins, M. L., B. Irvine, D. Tyner, E. Fine, c. Zayati, C. Chang, T. Horn, D. Ahle, J. Detmer, L. P. Shen, J. Kolberg, S. Bushnell, M. S. Urdea, and D. Ho. 1997. A branched DNA signal amplification assay for quantitation of nucleic acid targets below 100 molecules/ml. Nucleic Acids Res. 25:2979-2984. - PMC - PubMed
-
- Holguín, A., B. Rodés, U. Dietrich, and V. Soriano. 1999. HIV type 1 subtypes circulating in Spain. J. Med. Virol. 59:189-193. - PubMed
-
- Holguin, A., B. Rodés, and V. Soriano. 2000. Protease gene analysis of HIV type 1 non-B subtypes in Spain. AIDS Res. Hum. Retrovir. 16:1395-1403. - PubMed
-
- Jenny-Avital, E., and S. Beatrice. 2001. Erroneously low or undetectable plasma HIV-1 RNA load, determined by PCR, in West African and American patients with non-B subtype HIV-1 infection. Clin. Infect. Dis. 32:1227-1230. - PubMed