Comparative Study
. 2002 Apr;123(4):768-76.
doi: 10.1067/mtc.2002.121157.

Percutaneous aortic valve replacement: an experimental study. I. Studies on implantation

Affiliations
Free article
Comparative Study

Percutaneous aortic valve replacement: an experimental study. I. Studies on implantation

Georg Lutter et al. J Thorac Cardiovasc Surg. 2002 Apr.
Free article

Abstract

Objective: The purpose of this preliminary study was to devise a new surgical procedure for minimally invasive aortic valve implantation with a transluminal technique.

Methods: The new collapsible heart valve was prepared by mounting a porcine aortic valve, taken from a freshly slaughtered pig, into a self-expandable nitinol stent by means of a suture technique. The outer diameter of the valved stent ranged from 15 to 23 mm, and the length ranged from 21 to 28 mm. Before implantation in vivo, these valved stents were tested in an in vitro circulatory system. Only in vitro-tested valved stents with a pressure gradient of less than 7 mm Hg and regurgitation of I degrees or less were used for transluminal aortic valve implantation in vivo. Six of these valved stents were implanted in the descending aorta and 8 in the ascending aorta of anesthetized pigs. The catheter delivery system (22F) was extraperitoneally inserted through the left iliac artery or the infrarenal aorta. Measurements for transvalvular gradient, valvular opening and closure, blood-flow characteristics, regurgitation, and macroscopic analysis were performed at baseline and after the observation period (164 +/- 48 minutes).

Results: This preliminary study contained 14 animals. One animal died of ventricular fibrillation. Technical failure occurred in 2 pigs as a result of stent twisting. At the end of the observation period, the 11 successfully implanted valved stents demonstrated low transvalvular gradients (mean end-systolic Deltarho(max) of 5.4 +/- 3.3 mm Hg for the descending aorta group, 5.4 +/- 1.2 mm Hg for the supracoronary group, and 5.4 +/- 1.1 mm Hg for the subcoronary group), which did not differ from their in vitro gradients. Two-dimensional echocardiography demonstrated complete valvular closure and opening in 5 of 5 cases. Angiography indicated only a physiologic jet of regurgitation (0 degrees ) in 8 animals and mild (I degrees ) regurgitation in 3 animals. Color Doppler ultrasonography indicated no regurgitation in 5 of 5 cases and minor paravalvular leakage in 1 case.

Conclusion: Aortic valved stents can be successfully implanted without thoracotomy by using a transluminal catheter technique. Long-term function of the valves remains to be established.