Survival and polyethylene wear of porous-coated acetabular components in patients less than fifty years old: results at nine to fourteen years
- PMID: 12004013
- DOI: 10.2106/00004623-200205000-00005
Survival and polyethylene wear of porous-coated acetabular components in patients less than fifty years old: results at nine to fourteen years
Abstract
Background: Younger patients (those who are less than fifty years old) have been shown to have a high rate of failure of cemented acetabular components following total hip arthroplasty. In this report, we present the results associated with the use of an uncemented acetabular component in young patients who were evaluated at a minimum of nine years postoperatively.
Methods: Between December 1984 and December 1989, the senior author performed 174 primary total hip arthroplasties with use of a single design of porous-coated acetabular component. Seventy-one of these procedures were performed in fifty-six patients who were younger than fifty years old. Fifty-six of the seventy-one hips were available for radiographic and clinical analysis after a mean duration of follow-up of eleven years. All hips had been treated with a Harris-Galante-I porous-coated acetabular component that had been placed with a line-to-line fit and fixed with a mean of four screws. Clinical analysis was performed with use of the Harris hip score. Standardized anteroposterior radiographs were analyzed with regard to migration, radiolucent lines, pelvic osteolysis, and two-dimensional linear wear of the polyethylene.
Results: No metal shell was revised because of aseptic loosening, and no shell was loose at the time of the latest follow-up. A nonprogressive radiolucent line was seen in one zone in ten hips (18%) and in two zones in six hips (11%). No hip had a radiolucent line in all three zones. Pelvic osteolysis was noted in thirteen hips (23%); the osteolysis was observed in the ischium in eleven hips and around the screws in two. Survivorship analysis revealed that the probability of survival of the metal shell was 98% (95% confidence interval, 96.9% to 99.9%) at ten years. The mean rate of linear polyethylene wear (and standard deviation) was 0.15 +/- 0.10 mm/yr (range, 0.02 to 0.59 mm/yr). The wear rate was significantly increased in patients with an excellent Harris hip score (p = 0.004) and a younger age (less than thirty-eight years) (p = 0.026). With the numbers available, no relationship could be detected between the wear rate and the gender or weight of the patient, the polyethylene thickness, the abduction angle, or the femoral neck length.
Conclusions: The fixation and survival of porous-coated acetabular metal shells in patients less than fifty years old was excellent after a mean duration of follow-up of eleven years. The high rate of linear polyethylene wear and the high prevalence of pelvic osteolysis are of serious concern in this patient population. Continued follow-up will be necessary to evaluate the influence of these findings on the longevity of the fixation of this prosthesis.
Comment in
-
The need to account for patient activity when evaluating the results of total hip arthroplasty with survivorship analysis.J Bone Joint Surg Am. 2002 May;84(5):709-10. doi: 10.2106/00004623-200205000-00001. J Bone Joint Surg Am. 2002. PMID: 12004009 No abstract available.
Similar articles
-
The Harris-Galante porous-coated acetabular component with screw fixation. An average ten-year follow-up study.J Bone Joint Surg Am. 1999 Jan;81(1):66-73. doi: 10.2106/00004623-199901000-00010. J Bone Joint Surg Am. 1999. PMID: 9973056
-
Total hip replacement with a cementless acetabular component and a cemented femoral component in patients younger than fifty years of age.J Bone Joint Surg Am. 2002 May;84(5):770-4. doi: 10.2106/00004623-200205000-00011. J Bone Joint Surg Am. 2002. PMID: 12004019
-
Porous-coated acetabular components with screw fixation. Five to ten-year results.J Bone Joint Surg Am. 1996 Jul;78(7):975-81. doi: 10.2106/00004623-199607000-00001. J Bone Joint Surg Am. 1996. PMID: 8698733
-
High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty: a review of the PCA, the Harris Galante and the Profile/Tri-Lock Plus arthroplasties with a minimum of 12 years median follow-up in 96 hips.Acta Orthop. 2006 Aug;77(4):575-84. doi: 10.1080/17453670610012638. Acta Orthop. 2006. PMID: 16929433 Review.
-
Surgical management of polyethylene wear and pelvic osteolysis with modular uncemented acetabular components.J Arthroplasty. 2004 Jun;19(4 Suppl 1):124-9. doi: 10.1016/j.arth.2004.02.019. J Arthroplasty. 2004. PMID: 15190567 Review.
Cited by
-
Ceramic on ceramic hip arthroplasty in fused hips.Indian J Orthop. 2015 May-Jun;49(3):336-41. doi: 10.4103/0019-5413.156209. Indian J Orthop. 2015. PMID: 26015635 Free PMC article.
-
Is routine mid-term total hip arthroplasty surveillance beneficial?Clin Orthop Relat Res. 2012 Nov;470(11):3220-6. doi: 10.1007/s11999-012-2411-7. Epub 2012 Jun 6. Clin Orthop Relat Res. 2012. PMID: 22669552 Free PMC article.
-
THA using metal-on-metal articulation in active patients younger than 50 years.Clin Orthop Relat Res. 2008 Feb;466(2):340-6. doi: 10.1007/s11999-007-0045-y. Epub 2008 Jan 10. Clin Orthop Relat Res. 2008. PMID: 18196415 Free PMC article.
-
An unusual presentation of catastrophic failure of hip arthroplasty with a thigh mass.Arthroplast Today. 2016 Apr 8;2(2):63-67. doi: 10.1016/j.artd.2016.03.001. eCollection 2016 Jun. Arthroplast Today. 2016. PMID: 28326401 Free PMC article.
-
Hip resurfacing: a technology reborn.Postgrad Med J. 2006 Dec;82(974):802-5. doi: 10.1136/pgmj.2005.044594. Postgrad Med J. 2006. PMID: 17148702 Free PMC article. Review.