Evaluation of accuracy of computer-aided intraoperative positioning of endosseous oral implants in the edentulous mandible
- PMID: 12005146
- DOI: 10.1034/j.1600-0501.2002.130107.x
Evaluation of accuracy of computer-aided intraoperative positioning of endosseous oral implants in the edentulous mandible
Abstract
The overall accuracy of a novel surgical computer-aided navigation system for placement of endosseous implants was evaluated. Five dry cadaver mandibles were scanned using high resolution computed tomography (HRCT). The position of four interforaminal dental implants was planned on the computer screen and transferred to the cadaver mandibles using VISIT, a surgical navigation software developed at the Vienna General Hospital. The specimens were HRCT-scanned again to compare the position of the implants with the preoperative plan on reformatted slices after matching of the pre- and postoperative data sets using the mutual information technique. The overall accuracy was 0.96 +/- 0.72 mm (range 0.0-3.5 mm). No perforation of the mandibular cortex or damage to the mandibular canal occurred. We conclude that computer-aided implant surgery can reach a level of accuracy where further clinical developments are feasible.
Similar articles
-
Flapless implant surgery in the edentulous jaw based on three fixed intraoral reference points and image-guided surgical templates: accuracy in human cadavers.Clin Oral Implants Res. 2010 Aug;21(8):835-41. doi: 10.1111/j.1600-0501.2009.01884.x. Epub 2010 Mar 25. Clin Oral Implants Res. 2010. PMID: 20345381
-
Evaluation of the accuracy of three different computer-aided surgery systems in dental implantology: optical tracking vs. stereolithographic splint systems.Clin Oral Implants Res. 2008 Jul;19(7):709-16. doi: 10.1111/j.1600-0501.2007.01430.x. Epub 2008 May 19. Clin Oral Implants Res. 2008. PMID: 18492085
-
Accuracy of virtually planned and conventionally placed implants in edentulous cadaver maxillae and mandibles: a preliminary report.J Prosthet Dent. 2014 Oct;112(4):798-804. doi: 10.1016/j.prosdent.2014.05.008. Epub 2014 Jun 30. J Prosthet Dent. 2014. PMID: 24993376
-
Guided surgery for implant therapy.Oral Maxillofac Surg Clin North Am. 2011 May;23(2):239-56, v-vi. doi: 10.1016/j.coms.2011.01.008. Oral Maxillofac Surg Clin North Am. 2011. PMID: 21492799 Review.
-
Computerized Navigation for Minimal Invasive Retrieval of Displaced Dental Implants in the Lower Jaw.Oral Maxillofac Surg Clin North Am. 2019 Aug;31(3):447-455. doi: 10.1016/j.coms.2019.03.007. Epub 2019 May 17. Oral Maxillofac Surg Clin North Am. 2019. PMID: 31109843 Review.
Cited by
-
Effects of Crest Morphology on Lingual Concavity in Mandibular Molar Region: an Observational Study.J Oral Maxillofac Res. 2023 Mar 31;14(1):e3. doi: 10.5037/jomr.2023.14103. eCollection 2023 Jan-Mar. J Oral Maxillofac Res. 2023. PMID: 37180407 Free PMC article.
-
Guided implant surgery on oral cancer patients: in vitro study.Int J Comput Assist Radiol Surg. 2011 Jan;6(1):135-42. doi: 10.1007/s11548-010-0490-z. Epub 2010 Jul 27. Int J Comput Assist Radiol Surg. 2011. PMID: 20661657
-
Cone Beam CT Assessment of Bone Width of Upper and Lower Jaws for Dental Implant Placement: An Iraqi Study.Scientifica (Cairo). 2023 Apr 13;2023:4472154. doi: 10.1155/2023/4472154. eCollection 2023. Scientifica (Cairo). 2023. PMID: 37091876 Free PMC article.
-
Dental implant navigation system guide the surgery future.Kaohsiung J Med Sci. 2018 Jan;34(1):56-64. doi: 10.1016/j.kjms.2017.08.011. Epub 2017 Dec 8. Kaohsiung J Med Sci. 2018. PMID: 29310817 Free PMC article.
-
Accuracy of Dynamic Computer-Assisted Implant Placement: A Systematic Review and Meta-Analysis of Clinical and In Vitro Studies.J Clin Med. 2021 Feb 11;10(4):704. doi: 10.3390/jcm10040704. J Clin Med. 2021. PMID: 33670136 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources