In vivo confocal microscopy through-focusing to measure corneal flap thickness after laser in situ keratomileusis
- PMID: 12036637
- DOI: 10.1016/s0886-3350(02)01275-0
In vivo confocal microscopy through-focusing to measure corneal flap thickness after laser in situ keratomileusis
Abstract
Purpose: To measure flap thickness in laser in situ keratomileusis (LASIK) patients using in vivo confocal microscopy through-focusing (CMTF) and compare measured versus intended flap thickness achieved by 2 microkeratomes, the Automated Corneal Shaper(R) (ACS) (Chiron Bausch & Lomb) and the Hansatome (Bausch & Lomb).
Setting: Department of Ophthalmology, University of Texas Southwestern Medical Center at Dallas, Dallas, Texas, USA.
Methods: Twenty-seven eyes of 27 patients were examined by in vivo CMTF 3 to 12 months after LASIK was performed with the ACS (12 patients) or Hansatome (15 patients) microkeratome. The central cornea was scanned, and the epithelium, flap, stroma, and total corneal thickness were measured. Normalized flap thickness (NFT) was also calculated to account for changes in epithelial thickness that may have occurred postoperatively.
Results: The mean posterior stromal thickness was 341.1 microm +/- 53.9 (SD) (range 233 to 431 microm) in the ACS group and 320.3 +/- 42.3 microm (range 258 to 382 microm) in the Hansatome group. The mean nonnormalized flap thickness was 132.7 +/- 12.5 microm (range 11 to 151 microm) in the ACS group and 167.4 +/- 21.4 microm (range 141 to 209 microm) in the Hansatome group. The NFT was 129.6 +/- 9.5 microm and 158.4 +/- 22.1 microm, respectively. Both microkeratomes cut significantly less than intended (P <.05); however, the ACS cut a thinner-than-intended thickness in all cases, and the Hansatome cut thicker than intended in 13% of cases. The Hansatome also showed significantly greater variability in flap thickness than the ACS (P <.05).
Conclusions: A significant difference in precision was noted between the 2 microkeratomes. The findings emphasize the importance of performing thickness measurements and the usefulness of in vivo CMTF in making these determinations to ensure the safety and effectiveness of LASIK.
Similar articles
-
Comparison of corneal flap thickness between primary and fellow eyes using three microkeratomes.J Refract Surg. 2004 Sep-Oct;20(5):417-21. doi: 10.3928/1081-597X-20040901-02. J Refract Surg. 2004. PMID: 15523951 Clinical Trial.
-
Quality assessment model of 3 different microkeratomes through confocal microscopy.J Cataract Refract Surg. 2004 Jun;30(6):1300-9. doi: 10.1016/j.jcrs.2003.10.024. J Cataract Refract Surg. 2004. PMID: 15177608 Clinical Trial.
-
Laser in situ keratomileusis flap thickness using the Hansatome microkeratome with zero compression heads.J Cataract Refract Surg. 2006 Jan;32(1):72-7. doi: 10.1016/j.jcrs.2005.07.042. J Cataract Refract Surg. 2006. PMID: 16516782 Clinical Trial.
-
Flap and interface complications in LASIK.Curr Opin Ophthalmol. 2002 Aug;13(4):242-5. doi: 10.1097/00055735-200208000-00010. Curr Opin Ophthalmol. 2002. PMID: 12165708 Review.
-
Flap Thickness and the Risk of Complications in Mechanical Microkeratome and Femtosecond Laser In Situ Keratomileusis: A Literature Review and Statistical Analysis.Diagnostics (Basel). 2021 Aug 31;11(9):1588. doi: 10.3390/diagnostics11091588. Diagnostics (Basel). 2021. PMID: 34573930 Free PMC article. Review.
Cited by
-
Stromal thickness in the normal cornea: three-dimensional display with artemis very high-frequency digital ultrasound.J Refract Surg. 2009 Sep;25(9):776-86. doi: 10.3928/1081597X-20090813-04. Epub 2009 Sep 11. J Refract Surg. 2009. PMID: 19772263 Free PMC article.
-
In vivo confocal microscopy of the human cornea.Br J Ophthalmol. 2003 Feb;87(2):225-36. doi: 10.1136/bjo.87.2.225. Br J Ophthalmol. 2003. PMID: 12543757 Free PMC article. Review.
-
[Femtosecond laser for refractive corneal surgery: foundations, mode of action and clinical applications].Ophthalmologe. 2006 Dec;103(12):1005-13. doi: 10.1007/s00347-006-1450-4. Ophthalmologe. 2006. PMID: 17096101 Review. German.
-
Repeatability of laser in situ keratomileusis flap thickness measurement by Fourier-domain optical coherence tomography.J Cataract Refract Surg. 2011 Apr;37(4):649-54. doi: 10.1016/j.jcrs.2010.10.047. J Cataract Refract Surg. 2011. PMID: 21420588 Free PMC article.
-
Effect of Intraoperative Corneal Stromal Pocket Irrigation in Small Incision Lenticule Extraction.Biomed Res Int. 2015;2015:928608. doi: 10.1155/2015/928608. Epub 2015 Jul 27. Biomed Res Int. 2015. PMID: 26273659 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources