[Intraoperative electrophysiological monitoring of the recurrent laryngeal nerve in thyroid gland surgery--a prospective study]

[Article in German]
Affiliations
Clinical Trial

[Intraoperative electrophysiological monitoring of the recurrent laryngeal nerve in thyroid gland surgery--a prospective study]

[Article in German]
Th Friedrich et al. Zentralbl Chir. 2002 May.

Abstract

Introduction: In the last few years the use of intraoperative electrophysiological monitoring of the recurrent laryngeal nerve (RLN) in thyroid gland surgery has become more and more important.

Patients and method: In a prospective study 223 nerves at risk in 116 patients were monitored with the Neurosign(R)100 (Fa. Magstim Ltd., UK). We used intramuscular needle electrodes inserted into the vocal muscle through the conic ligament. Practicability, complications, acceptance and predictive value of the method were documented. Recurrent nerve palsy rate and complications were compared with a control group operated upon without monitoring.

Results: The intraoperative delay using this method was on average 8.9 minutes. There were problems with monitoring equipment avoiding use in 6.4 %. In 2 cases (1.7 %) an accidental lesion of endotracheal tube cuff was found related to malpositioning of the needle and in 7.7 % a hematoma of the vocal cords was observed. 73.3 % of the surgeons accepted the method to identify and control the nerve integrity. False-positive and false-negative signals may occur. In cases of a final real stimulus response a regular vocal cord motility was found in 95 %. If a nerve conduction block was noted an immobility of ipsilateral vocal cord was diagnosed postoperatively in 50 %. There was no decrease in transient recurrent palsy rate using monitoring (10.7 % vs. 9.6 % without monitoring) but in permanent paralysis (1.8 % vs. 3.0 %).

Conclusions: It may be concluded that intraoperative electrophysiological monitoring of the RLN is a simple and accepted method with low complications reducing the incidence of permanent RLN palsy rate. We found the monitoring especially useful for operations of recurrent goiter and carcinomas of the thyroid gland as well as for learning thyroid gland surgery.