A reusable, custom-made warming blanket prevents core hypothermia during major neonatal surgery
- PMID: 12067874
- DOI: 10.1007/BF03017389
A reusable, custom-made warming blanket prevents core hypothermia during major neonatal surgery
Abstract
Purpose: To introduce a reusable model of neonatal forced air warming blanket for intraoperative use during major noncardiac neonatal surgery and to determine clinical efficacy of this reusable blanket compared with the commonly used disposable blankets.
Methods: Delivered air temperature and calorie uptake of standard thermal bodies within the reusable blankets, Bair Hugger(R) blanket model 530 and model 555 were studied. Also, an efficacy study was conducted in 90 neonatal patients scheduled for major noncardiac surgery comparing the reusable blanket, the Bair Hugger(R) blanket model 530 and passive heat conservation as a control. The covered reusable blanket was used as a rescue procedure if the core temperature was < 35.5 degrees C.
Results: Delivered air temperature and heat transfer from the covered reusable blanket did not differ significantly from those of the Bair Hugger(R) blanket model 530 and model 555 (despite 0.75 degrees C-1.2 degrees C of heat trapped under the sheet and 1.3 Kcal less energy transfer). Temperatures measured underneath patients (correlated to poorly perfused areas) were highest using the Bair Hugger(R) blanket model 555. The reusable blanket was efficacious in preventing intraoperative core hypothermia and not different from the Bair Hugger(R) blanket model 530. About 1/3 of the patients in the control group had presented a core temperature < 35.5 degrees C but were successfully rescued using the reusable blanket. No adverse events were associated with any of these warming methods.
Conclusion: This study shows the clinical efficacy of our reusable blanket for the prevention of core hypothermia during major neonatal surgery, which is not different from commonly used disposable blankets.
Similar articles
-
Comparison of forced-air warming systems with upper body blankets using a copper manikin of the human body.Acta Anaesthesiol Scand. 2002 Sep;46(8):965-72. doi: 10.1034/j.1399-6576.2002.460807.x. Acta Anaesthesiol Scand. 2002. PMID: 12190797
-
Comparison of forced-air warming systems with lower body blankets using a copper manikin of the human body.Acta Anaesthesiol Scand. 2003 Jan;47(1):58-64. doi: 10.1034/j.1399-6576.2003.470110.x. Acta Anaesthesiol Scand. 2003. PMID: 12492798
-
Comparison of two convective warming systems during major abdominal and orthopedic surgery.Can J Anaesth. 2008 Jun;55(6):358-63. doi: 10.1007/BF03021491. Can J Anaesth. 2008. PMID: 18566199 Clinical Trial.
-
Forced-air warming: technology, physical background and practical aspects.Curr Opin Anaesthesiol. 2009 Dec;22(6):769-74. doi: 10.1097/ACO.0b013e328331d134. Curr Opin Anaesthesiol. 2009. PMID: 19734783 Review.
-
Prevention of perioperative hypothermia with forced-air warming systems and upper-body blankets.Surg Technol Int. 2006;15:19-22. Surg Technol Int. 2006. PMID: 17029156 Review.
Cited by
-
Active body surface warming systems for preventing complications caused by inadvertent perioperative hypothermia in adults.Cochrane Database Syst Rev. 2016 Apr 21;4(4):CD009016. doi: 10.1002/14651858.CD009016.pub2. Cochrane Database Syst Rev. 2016. PMID: 27098439 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical