Preliminary experience with the St. Jude Medical Regent mechanical heart valve in the aortic position: early in vivo hemodynamic results

Affiliations

Preliminary experience with the St. Jude Medical Regent mechanical heart valve in the aortic position: early in vivo hemodynamic results

Sandro Gelsomino et al. Ann Thorac Surg. 2002 Jun.

Abstract

Background: The St. Jude Medical Regent is a new generation mechanical aortic valve.

Methods: Between March 2000 and July 2001, this valve was implanted in the aortic position in 40 patients (21 men; mean age 59.1 +/- 9.0 years). Preoperatively, 24 patients (60%) were in New York Heart Association functional class III or IV. Eighteen patients (45%) underwent associated procedures. Mean valve size was 21.4 +/- 2.4 mm. The mean duration of follow-up was 8.5 +/- 4.5 months (range, 1 to 16 months).

Results: There were no operative deaths. Early complications included one reoperation for bleeding and one transient low output syndrome. Valve replacement was followed by a significant reduction in mean and peak transaortic gradients over time (p < 0.001) and analysis of variance failed to demonstrate statistical differences between valve size over time (p = not significant). A significant reduction in left ventricular hypertrophy occurred over time (p = 0.01) in all valve sizes (p = not significant between groups): baseline left ventricular mass index was 194 g/cm2; it reduced by 22 g/cm2 (p = 0.006) at discharge. Left ventricular mass index decreased from 172 +/- 55 g/cm2 to 156 +/- 44 g/cm2 (p = 0.03) from discharge to 2 months. Further reductions were not significant. Relative wall thickness decreased from 0.57 +/- 0.13 preoperatively to 0.42 +/- 0.06 at discharge (p = 0.001), and again at 2 months (-0.2; p = not significant), and at 1 year (-0.02; p = not significant).

Conclusions: The early experience with the St. Jude Medical Regent valve has been satisfactory.