Morphological predictors of restenosis after coronary stenting in humans
- PMID: 12081990
- DOI: 10.1161/01.cir.0000019071.72887.bd
Morphological predictors of restenosis after coronary stenting in humans
Abstract
Background: Experimental studies suggest that arterial injury and inflammation lead to increased neointimal growth after stenting. Despite the increased use of coronary stents in humans, there are only limited pathological data on the morphological features of in-stent restenosis.
Methods and results: Detailed histology was performed on 116 stents, implanted > or =90 days in 87 coronary arteries, from 56 patients (mean age, 59+/-13 years). The mean duration of stent implant was 10 months. In-stent restenosis was defined as a stent area stenosis of >75%. Lumen area increased as stent area increased (r2=0.27, P=0.0001), but there was a much stronger correlation between stent area and neointimal area (r2=0.70, P<0.0001). Arterial medial fracture was associated with a 29% increase (P<0.01) in neointimal thickness compared with arteries with an intact media. Neointimal thickness (P=0.0001), inflammatory cell density (P<0.0001), and neointimal vascular channel density (P<0.0001) were greater when struts were in contact with a ruptured arterial media compared with fibrous plaque or an intact fibrous cap. Stent strut penetration into a lipid core was associated with increased neointimal thickness (P=0.04) and inflammatory cell density (P=0.03). Neointimal inflammatory cell content was 2.4-fold greater in stents with restenosis versus no restenosis, and inflammation was associated with increased neoangiogenesis.
Conclusions: Coronary stenting that is accompanied by medial damage or penetration of the stent into a lipid core induces increased arterial inflammation, which is associated with increased neointimal growth. These data suggest the use of stenting strategies that reduce inflammation and neoangiogenesis to reduce the incidence of restenosis.
Comment in
-
What is "the matter" with restenosis in 2002?Circulation. 2002 Jun 25;105(25):2932-3. doi: 10.1161/01.cir.0000023396.71735.65. Circulation. 2002. PMID: 12081980 No abstract available.
Similar articles
-
Pathology of acute and chronic coronary stenting in humans.Circulation. 1999 Jan 5-12;99(1):44-52. doi: 10.1161/01.cir.99.1.44. Circulation. 1999. PMID: 9884378
-
In-stent restenosis is associated with neointimal angiogenesis and macrophage infiltrates.Pathol Res Pract. 2014 Dec;210(12):1026-30. doi: 10.1016/j.prp.2014.04.004. Epub 2014 Apr 18. Pathol Res Pract. 2014. PMID: 24815534
-
Peri-strut low-intensity areas in optical coherence tomography correlate with peri-strut inflammation and neointimal proliferation: an in-vivo correlation study in the familial hypercholesterolemic coronary swine model of in-stent restenosis.Coron Artery Dis. 2014 Nov;25(7):595-601. doi: 10.1097/MCA.0000000000000134. Coron Artery Dis. 2014. PMID: 25013949
-
Comparative pathology: radiation-induced coronary artery disease in man and animals.Semin Interv Cardiol. 1998 Sep-Dec;3(3-4):163-72. Semin Interv Cardiol. 1998. PMID: 10406688 Review.
-
Prevention of restenosis after stenting: the emerging role of inflammation.Coron Artery Dis. 2004 Sep;15(6):307-11. doi: 10.1097/00019501-200409000-00002. Coron Artery Dis. 2004. PMID: 15346087 Review.
Cited by
-
Clinical outcome of biodegradable polymer sirolimus-eluting stent and durable polymer everolimus-eluting stent in patients with diabetes.Cardiovasc Diabetol. 2020 Oct 1;19(1):162. doi: 10.1186/s12933-020-01145-x. Cardiovasc Diabetol. 2020. PMID: 33004019 Free PMC article.
-
ON VASCULAR STENOSIS, RESTENOSIS AND MANNOSE BINDING LECTIN.Arq Bras Cir Dig. 2016 Mar;29(1):57-9. doi: 10.1590/0102-6720201600010015. Arq Bras Cir Dig. 2016. PMID: 27120743 Free PMC article. Review.
-
Systematical evolution on a Zn-Mg alloy potentially developed for biodegradable cardiovascular stents.J Mater Sci Mater Med. 2019 Nov 1;30(11):122. doi: 10.1007/s10856-019-6324-9. J Mater Sci Mater Med. 2019. PMID: 31677119
-