Continuous blood gas monitoring using an in-dwelling optode method: clinical evaluation of the Neotrend sensor using a luer stub adaptor to access the umbilical artery catheter
- PMID: 12082470
- DOI: 10.1038/sj.jp.7210735
Continuous blood gas monitoring using an in-dwelling optode method: clinical evaluation of the Neotrend sensor using a luer stub adaptor to access the umbilical artery catheter
Abstract
Introduction: Arterial blood gases are essential in the management of critically ill neonates. A new technology using the Neotrend system (Diametrics Medical) allows for continuous measurement of pH, PaCO(2), and PaO(2), and calculates oxygen saturation, bicarbonate, and base excess.
Objective: To evaluate and compare the results of continuous blood gas measurement using the Neotrend system with a standard system of blood gas analysis in our intensive care unit.
Design: Prospective, controlled, interventional study.
Setting: The neonatal intensive care unit of a tertiary referral center.
Patients: Neonates with respiratory distress who required respiratory support and frequent arterial blood gas sampling and had a UAC.
Research design and method: To enable Neotrend sensor access to an existing Argyle umbilical artery catheter (UAC) the catheter was cut at the 25-cm mark and connected to an 18-gauge blunt needle luer stub adaptor (Vygon 95440). The study began with the insertion of the Neotrend sensor. Subjects remained on the study until the UAC was discontinued and/or frequent blood gases were no longer needed. The blood gas results from the Neotrend system were not used in clinical management of the patient. BLOOD GAS MEASUREMENT: During the study period, with each blood sample drawn for laboratory analysis, a printout from the Neotrend monitor was recorded for comparison.
Results: A total of 217 pairs of blood gas samples were collected from seven neonates. The mean bias/precision for pH was 0.01/0.04; for PaO(2) 0.72/18.5 mm Hg; and for PaCO(2) 3.96/2.63 mm Hg. The correlation (r value) between the sensor reading and the blood gases were 0.85 for pH, 0.96 for PaO(2), and 0.92 for PaCO(2).
Conclusion: The blood gases compared in the two methods had a strong correlation for pH, PaCO(2), PaO(2), and oxygen saturation. Although the bicarbonate and base excess values showed suboptimal statistical correlation, the difference was not clinically relevant. Results of this study indicate that this technology provides an accurate means of monitoring continuous blood gas parameters in neonatal patients. It also allows reduced healthcare provider exposure to blood and decreased patient iatrogenic blood loss.
Similar articles
-
Continuous blood gas monitoring using an in-dwelling optode method: comparison to intermittent arterial blood gas sampling in ECMO patients.J Perinatol. 2002 Sep;22(6):472-4. doi: 10.1038/sj.jp.7210776. J Perinatol. 2002. PMID: 12168125 Clinical Trial.
-
Clinical validation of a continuous intravascular neonatal blood gas sensor introduced through an umbilical artery catheter.Respir Care. 2002 Jun;47(6):682-7. Respir Care. 2002. PMID: 12036438
-
Evaluation of a new combined transcutaneous measurement of PCO2/pulse oximetry oxygen saturation ear sensor in newborn patients.Pediatrics. 2005 Jan;115(1):e64-8. doi: 10.1542/peds.2004-0946. Epub 2004 Dec 15. Pediatrics. 2005. PMID: 15601814
-
[Non-invasive method of arterial oxygen saturation monitoring in the neonatal intensive care unit (with a review of blood oxygen monitoring methods)].Tijdschr Kindergeneeskd. 1988 Feb;56(1):20-6. Tijdschr Kindergeneeskd. 1988. PMID: 3282343 Review. Dutch.
-
Emerging technology in critical care: continuous intra-arterial blood gas monitoring.Am J Crit Care. 1996 Jan;5(1):55-65. Am J Crit Care. 1996. PMID: 8680494 Review.
MeSH terms
LinkOut - more resources
Full Text Sources