Two-dimensional spatially-selective RF excitation pulses in echo-planar imaging
- PMID: 12111965
- DOI: 10.1002/mrm.10157
Two-dimensional spatially-selective RF excitation pulses in echo-planar imaging
Abstract
Two-dimensional spatially-selective RF (2DRF) excitation pulses were developed for single-shot echo-planar imaging (EPI) with reduced field of view (FOV) in the phase-encoding direction. The decreased number of k-space lines significantly shortens the length of the EPI echo train. Thus, both gradient-echo and spin-echo 2DRF-EPI images of the human brain at 2.0 T exhibit markedly reduced susceptibility artifacts in regions close to major air cavities. Based on a blipped-planar trajectory, implementation of a typical 2DRF pulse resulted in a 26-ms pulse duration, a 5-mm section thickness, a 40-mm FOV along the phase-encoding direction, and a 200-mm distance of the unavoidable side excitations from the center of the FOV. For the above conditions and at 2 x 2 mm(2) resolution, 2DRF-EPI yielded an echo train length of only 21 ms, as opposed to 102 ms for conventional EPI. This gain in time may be used to achieve higher spatial resolution. For example, spin-echo 2DRF-EPI of a 40-mm FOV at 1 x 1 mm(2) resolution led to an echo train of 66 ms. Although the current implementation still lacks user-friendliness, 2DRF pulses are likely to become a useful addition to the arsenal of advanced MRI tools. .
Copyright 2002 Wiley-Liss, Inc.
Similar articles
-
Fast-spin-echo imaging of inner fields-of-view with 2D-selective RF excitations.J Magn Reson Imaging. 2010 Jun;31(6):1530-7. doi: 10.1002/jmri.22196. J Magn Reson Imaging. 2010. PMID: 20512911
-
Functional neuroimaging of inner fields-of-view with 2D-selective RF excitations.Magn Reson Imaging. 2013 Sep;31(7):1228-35. doi: 10.1016/j.mri.2013.03.005. Epub 2013 Apr 18. Magn Reson Imaging. 2013. PMID: 23602726
-
Gradient-echo line scan imaging using 2D-selective RF excitation.J Magn Reson. 2000 Nov;147(1):17-25. doi: 10.1006/jmre.2000.2159. J Magn Reson. 2000. PMID: 11042043
-
Small Field-of-view single-shot EPI-DWI of the prostate: Evaluation of spatially-tailored two-dimensional radiofrequency excitation pulses.Z Med Phys. 2016 Jun;26(2):168-76. doi: 10.1016/j.zemedi.2015.06.013. Epub 2015 Aug 20. Z Med Phys. 2016. PMID: 26300045
-
K-space in the clinic.J Magn Reson Imaging. 2004 Feb;19(2):145-59. doi: 10.1002/jmri.10451. J Magn Reson Imaging. 2004. PMID: 14745747 Review.
Cited by
-
Magnetic resonance imaging of the cerebellopontine angle: comparison between constructive interference steady-state and small field-of-view technique turbo spin echo sequences.Radiologie (Heidelb). 2024 Nov;64(Suppl 1):123-130. doi: 10.1007/s00117-024-01334-9. Epub 2024 Jul 10. Radiologie (Heidelb). 2024. PMID: 38985167 English.
-
Multi-parametric quantitative in vivo spinal cord MRI with unified signal readout and image denoising.Neuroimage. 2020 Aug 15;217:116884. doi: 10.1016/j.neuroimage.2020.116884. Epub 2020 Apr 29. Neuroimage. 2020. PMID: 32360689 Free PMC article.
-
Zoomed EPI-DWI of the head and neck with two-dimensional, spatially-selective radiofrequency excitation pulses.Eur Radiol. 2014 Oct;24(10):2507-12. doi: 10.1007/s00330-014-3287-6. Epub 2014 Jul 5. Eur Radiol. 2014. PMID: 24996796
-
Reduced field-of-view single-shot fast spin echo imaging using two-dimensional spatially selective radiofrequency pulses.J Magn Reson Imaging. 2010 Jul;32(1):242-8. doi: 10.1002/jmri.22204. J Magn Reson Imaging. 2010. PMID: 20578031 Free PMC article.
-
Simultaneous multi-segment (SMSeg) EPI over multiple focal regions.Phys Med Biol. 2023 Feb 3;68(4):10.1088/1361-6560/acb2a9. doi: 10.1088/1361-6560/acb2a9. Phys Med Biol. 2023. PMID: 36634366 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources