Diagnostic accuracy of stereotactic large-core needle biopsy for nonpalpable breast disease: results of a multicenter prospective study with 95% surgical confirmation
- PMID: 12115488
- DOI: 10.1002/ijc.10419
Diagnostic accuracy of stereotactic large-core needle biopsy for nonpalpable breast disease: results of a multicenter prospective study with 95% surgical confirmation
Abstract
Stereotactic large-core needle biopsy is increasingly applied for the diagnosis of nonpalpable breast disease. Our study examines whether this minimally invasive technique is sufficiently accurate to replace surgical breast biopsy. In a prospective multicenter study, 973 consecutive women with 1,029 nonpalpable breast lesions were offered stereotactic 14-gauge needle biopsy. If the needle biopsy yielded breast cancer, the patient was offered therapeutic surgery. Surgical biopsy was proposed in cases of needle biopsies without malignancy. An expert panel reviewed all discrepancies in histologic diagnosis between the needle biopsy and open biopsy. Forty-five patients withdrew from participation and 113 (11%) planned needle biopsy procedures were cancelled. Of the 871 successful biopsy procedures, 95% were confirmed surgically. In 13 cases (1.5%), insufficient material was obtained for histologic assessment. Fifty-five percent of the needle biopsies were diagnosed as malignant (290 invasive cancers, 190 ductal carcinoma in situ). Thirteen of the 322 lesions (4%, 95% CI 2-7%) with a benign needle biopsy diagnosis contained malignancy after surgery. Six of the 26 (23%, 95% CI 9-44%) lesions with a high-risk diagnosis (atypical ductal or lobular hyperplasia or lobular carcinoma in situ) were diagnosed as malignant after surgery. Five of the 30 lesions containing normal breast tissue held malignancy (17%, 95% CI 6-35%). Guidelines for the management of different categories of needle biopsy diagnoses were made. Application of these guidelines to the present findings resulted in sensitivity and specificity rates of 97% (95% CI 95-98%) and 99% (95% CI 97-100%), respectively. Stereotactic large-core needle biopsy is an accurate diagnostic instrument for nonpalpable breast disease. It may safely replace needle localised open-breast biopsy provided that high-risk and normal breast tissue diagnoses are followed by needle or open-breast biopsy.
Copyright 2002 Wiley-Liss, Inc.
Similar articles
-
[Breast calcifications with percutaneous vacuum-assisted biopsy diagnosis of malignancy or atypical hyerplasia: correlations with surgical findings].Pathologica. 2002 Dec;94(6):299-305. doi: 10.1007/s102420200053. Pathologica. 2002. PMID: 12540993 Italian.
-
Stereotactic and sonographic large-core biopsy of nonpalpable breast lesions: results of the Radiologic Diagnostic Oncology Group V study.Acad Radiol. 2004 Mar;11(3):293-308. doi: 10.1016/s1076-6332(03)00510-5. Acad Radiol. 2004. PMID: 15035520 Clinical Trial.
-
Upstaging of atypical ductal hyperplasia after vacuum-assisted 11-gauge stereotactic core needle biopsy.Arch Surg. 2003 Jun;138(6):619-22; discussion 622-3. doi: 10.1001/archsurg.138.6.619. Arch Surg. 2003. PMID: 12799332
-
Role of stereotactic breast biopsy.Semin Surg Oncol. 1996 Sep-Oct;12(5):290-9. doi: 10.1002/(SICI)1098-2388(199609/10)12:5<290::AID-SSU3>3.0.CO;2-8. Semin Surg Oncol. 1996. PMID: 8873316 Review.
-
Interventional breast procedures.Curr Probl Diagn Radiol. 1998 Sep-Oct;27(5):133-84. doi: 10.1016/s0363-0188(98)90010-x. Curr Probl Diagn Radiol. 1998. PMID: 9794095 Review.
Cited by
-
Mapping breast tissue types by miniature radio-frequency near-field spectroscopy sensor in ex-vivo freshly excised specimens.BMC Med Imaging. 2016 Oct 10;16(1):57. doi: 10.1186/s12880-016-0160-x. BMC Med Imaging. 2016. PMID: 27724884 Free PMC article.
-
Development and evaluation of a prediction model for underestimated invasive breast cancer in women with ductal carcinoma in situ at stereotactic large core needle biopsy.PLoS One. 2013 Oct 11;8(10):e77826. doi: 10.1371/journal.pone.0077826. eCollection 2013. PLoS One. 2013. PMID: 24147085 Free PMC article.
-
Trends in breast biopsies for abnormalities detected at screening mammography: a population-based study in the Netherlands.Br J Cancer. 2013 Jul 9;109(1):242-8. doi: 10.1038/bjc.2013.253. Epub 2013 May 21. Br J Cancer. 2013. PMID: 23695018 Free PMC article.
-
Surgical outcome of patients with core-biopsy-proven nonpalpable breast carcinoma: a large cohort follow-up study.Ann Surg Oncol. 2009 Aug;16(8):2252-8. doi: 10.1245/s10434-009-0513-6. Epub 2009 May 13. Ann Surg Oncol. 2009. PMID: 19437077 Free PMC article.
-
Analysis of stereotactic vacuum-assisted breast biopsy for patients with segmental calcifications.Jpn J Radiol. 2009 Dec;27(10):450-4. doi: 10.1007/s11604-009-0370-z. Epub 2009 Dec 25. Jpn J Radiol. 2009. PMID: 20035418
Publication types