Clinical Trial
. 2002 Jul;28(7):947-52.
doi: 10.1007/s00134-002-1334-2. Epub 2002 Jun 15.

Cardiac output determination in children: equivalence of the transpulmonary thermodilution method to the direct Fick principle

Affiliations
Clinical Trial

Cardiac output determination in children: equivalence of the transpulmonary thermodilution method to the direct Fick principle

C Pauli et al. Intensive Care Med. 2002 Jul.

Abstract

Objective: To show the equivalence of the transpulmonary thermodilution method to the direct Fick principle in children.

Design: Prospective single-centre study.

Setting: A 16-bed paediatric cardiac ICU and a cardiac catheterisation laboratory at an university affiliated centre for paediatric cardiology and congenital heart disease.

Patients: We consecutively investigated 18 patients (mean age 12.1 +/- 6.4 years) during cardiac catheterisation and after corrective cardiac operation.

Methods and results: We prospectively defined limits of equivalence for cardiac index (CI) for both methods of +/- 0.25 l/min x m(2). We measured oxygen consumption for determination of CI by Fick as the clinical "gold standard" and performed a set of three transpulmonary thermodilution measurements. The mean CI(Fick) was 2.88 +/- 1.07 l/min x m(2) (range 1.10-4.62 l/min x m(2)) and CI(TPID)was 2.85 +/- 1.03 l/min x m(2)(range 1.02-4.49 l/min x m(2)). The mean difference between CI(Fick) and CI(TPID)was 0.030 +/- 0.168 l/min x m(2), and limits of agreement -0.306 to 0.366 l/min x m(2)(90% confidence interval -0.040 to 0.099 l/min x m(2)). The regression equation was : CI(Fick)=1.0244 x CI(TPID)-0.040, r(2) = 0.976, P < 0.0001. The intraclass coefficient of reliability for three repeated measurements of CI(TPID) was 0.97, the corresponding lower limit of the 95% confidence interval was 0.94.

Conclusion: We demonstrated the equivalence of CI measurement by transpulmonary thermodilution and the Fick principle in children. This new method may improve hemodynamic monitoring and management in seriously ill children.

PubMed Disclaimer