Primary radiation outside the imaged volume of a multislice helical CT scan
- PMID: 12124239
- DOI: 10.1259/bjr.75.894.750518
Primary radiation outside the imaged volume of a multislice helical CT scan
Abstract
Multislice helical CT scanning has advantages of speed and X-ray tube loading, making it possible to image larger volumes in a single exposure. Our aim is to investigate dose implications for short scans from the additional X-ray tube rotations required to reconstruct a given volume in helical scanning. To this end a multislice scanner was compared with a single slice scanner. Two independent methods were used. The first was based on optical density measurements taken from a film exposed free-in-air as it moves with the CT bed along the scan axis. The second used measurements from a pencil ionization chamber supported free-in-air at the centre of the CT aperture for the duration of both a long scan and a short scan. This method assumes the same excess primary radiation at the extremes of both scans and the measurements are incorporated into two simultaneous equations. The dose-length product outside the imaged volume has been compared with the dose-length product inside the imaged volume using both methods. For 4 x 5 mm multislice collimation with a 360 degrees interpolation and a pitch of 0.875, the film and simultaneous equations methods show an excess dose-length product at the extremes of the scan volume equivalent to 3.3 cm and 3.5 cm extra scan length, respectively. This represents a large percentage of a short scan and is substantially greater than for a helical scan using the single slice scanner with 5 mm collimation, a 360 degrees interpolation and a pitch of 1. The latter showed an excess dose-length product at the extremes which was equivalent to 0.35 cm scan length by the film method and 0.25 cm using simultaneous equations. Taking the abdominal protocols recommended by the respective manufacturers, however, the multislice scanner could cover a 45 cm scan length in a single exposure, while the single slice scanner needed six exposures to image the same volume. With the multislice scanner set at 4 x 2.5 mm collimation, 360 degrees interpolation and a pitch of 0.875, the dose-length product outside the volume of interest was equivalent to 1.9 cm scan length by the first method and 1.8 cm using the second method. With 4 x 1 mm collimation it was equivalent to 1.0 cm using both methods. Changing the interpolation algorithm from 360 degrees to 180 degrees had no effect on the additional equivalent scan length while doubling the pitch resulted in a 25% increase. We conclude from this study that with the multislice scanner, the axial mode is to be preferred for short CT scans such as those used in patient biopsies and drainage. For paediatric helical scans shorter than 13 cm, dose length product is reduced by using 4 x 2.5 mm rather than 4 x 5 mm collimation. For longer scans, however, the increased CT dose index associated with narrower collimation in the multislice mode offsets the dose reduction at the extremes.
Similar articles
-
The impact of overscan on patient dose with first generation multislice CT scanners.Phys Med. 2011 Apr;27(2):69-74. doi: 10.1016/j.ejmp.2010.03.001. Epub 2010 Apr 24. Phys Med. 2011. PMID: 20456992
-
The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations.Med Phys. 2005 Jun;32(6):1621-9. doi: 10.1118/1.1924309. Med Phys. 2005. PMID: 16013721
-
Dose and pitch relationship for a particular multislice CT scanner.AJR Am J Roentgenol. 2001 Dec;177(6):1273-5. doi: 10.2214/ajr.177.6.1771273. AJR Am J Roentgenol. 2001. PMID: 11717063
-
Multislice CT: technical principles and future trends.Eur Radiol. 2003 Dec;13 Suppl 5:M3-13. doi: 10.1007/s00330-003-2178-z. Eur Radiol. 2003. PMID: 14989605 Review.
-
[Dental CT: image quality and radiation exposure in relation to scan parameters].Rofo. 1999 Feb;170(2):137-44. doi: 10.1055/s-2007-1011025. Rofo. 1999. PMID: 10101352 Review. German.
Cited by
-
A methodology for direct quantification of over-ranging length in helical computed tomography with real-time dosimetry.J Appl Clin Med Phys. 2011 Jan 31;12(2):3326. doi: 10.1120/jacmp.v12i2.3326. J Appl Clin Med Phys. 2011. PMID: 21587174 Free PMC article.
-
Open-Source Radiation Exposure Extraction Engine (RE3) with Patient-Specific Outlier Detection.J Digit Imaging. 2016 Aug;29(4):406-19. doi: 10.1007/s10278-015-9852-y. J Digit Imaging. 2016. PMID: 26644157 Free PMC article.
-
Reduction of the unnecessary dose from the over-range area with a spiral dynamic z-collimator: comparison of beam pitch and detector coverage with 128-detector row CT.Radiol Phys Technol. 2012 Jan;5(1):53-8. doi: 10.1007/s12194-011-0135-0. Epub 2011 Oct 15. Radiol Phys Technol. 2012. PMID: 22002584
-
A method for calculating the dose length product from CT DICOM images.Br J Radiol. 2011 Mar;84(999):236-43. doi: 10.1259/bjr/37630380. Epub 2010 Nov 16. Br J Radiol. 2011. PMID: 21081566 Free PMC article.
-
Patient dose considerations in computed tomography examinations.World J Radiol. 2010 Jul 28;2(7):262-8. doi: 10.4329/wjr.v2.i7.262. World J Radiol. 2010. PMID: 21160666 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical